SwePub
Sök i LIBRIS databas

  Extended search

onr:"swepub:oai:research.chalmers.se:ce3ff560-fc10-4aaf-a098-3c790d9b2cb7"
 

Search: onr:"swepub:oai:research.chalmers.se:ce3ff560-fc10-4aaf-a098-3c790d9b2cb7" > Mechanical load app...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Mechanical load applied by Intraosseous Transcutaneous Amputation Prosthesis (ITAP) during walking on level and sloped treadmill: A case study

Ahmed, Kirstin, 1974 (author)
University College London (UCL)
Thornton, Matthew (author)
University College London (UCL)
Taylor, Stephen (author)
University College London (UCL)
 (creator_code:org_t)
2023
2023
English.
In: Medical Engineering and Physics. - 1350-4533 .- 1873-4030.
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • This proof of concept study presents a method to collect and analyse kinetic data from one participant with a transfemoral amputation fitted with a percutaneous osseointegrated implant walking on a level and sloped treadmill. We describe the construction of and results from a bespoke wireless six axis load cell built into one participant’s prosthetic assembly. The load cell does not clinically compromise the participant in any way and is an initial milestone in the development of a light-weight wireless load cell for use with percutaneous osseointegrated implants. In this case, it is the first time that kinetic data from a participant fitted with an Intraosseous Transcutaneous Amputation Prosthesis has been published. We propose that the data can be used to model the load transfer to the host bone, with several clinically significant applications. The raw dynamic data are made available and quasi-static load cases for each functional phase of gait are presented. Peak forces obtained in the medio-lateral (X), cranio-caudal (Y) and antero-posterior (Z) axes over level ground respectively were -243.8 N (0.24 BW), 1321.5 N (1.31 BW) and -421.8 N (0.42 BW); uphill were -141.0 N (0.14 BW), 1604.2 N (1.59 BW), -498.1 (0.49 BW); downhill were -206.0 N (0.20 BW), 1103.9 N (1.09 BW), -547.2 N (0.54 BW). The kinetics broadly followed non-amputated gait patterns with some gait strategies consistent in participants with other implant designs or prosthetic socket connections, for example offloading the artificial limb downhill.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Hälsovetenskap -- Sjukgymnastik (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Health Sciences -- Physiotherapy (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Ahmed, Kirstin, ...
Thornton, Matthe ...
Taylor, Stephen
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Health Sciences
and Physiotherapy
ENGINEERING AND TECHNOLOGY
ENGINEERING AND ...
and Medical Engineer ...
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
and Orthopaedics
Articles in the publication
Medical Engineer ...
By the university
Chalmers University of Technology

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view