Brown, Shannon (författare)
-
- Ortiz Catalan, Max Jair, 1982 (författare)
- Chalmers University of Technology,Chalmers tekniska högskola
-
- Petersson, Joel (författare)
- Högskolan i Borås,Akademin för textil, teknik och ekonomi,University of Borås
-
visa fler...
-
- Rödby, Kristian (författare)
- Högskolan i Borås,Akademin för textil, teknik och ekonomi,University of Borås
-
- Seoane, Fernando, 1976- (författare)
- KTH,Högskolan i Borås,Akademin för textil, teknik och ekonomi,KTH-School of Technology and Health,Data- och elektroteknik,University of Borås, Sweden,Kungliga Tekniska Högskolan (KTH),Royal Institute of Technology (KTH)
-
- Brown, Stephen H. (författare)
- Högskolan i Borås,University of Borås
-
visa färre...
-
(creator_code:org_t)
- ISBN 9781457702204
- Institute of Electrical and Electronics Engineers (IEEE), 2016
- 2016
- Engelska.
-
Ingår i: Engineering in Medicine and Biology Society (EMBC), 2016 IEEE 38th Annual International Conference of the. - : Institute of Electrical and Electronics Engineers (IEEE). - 1557-170X. - 9781457702204 ; , s. 6074-6077
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
https://research.cha...
-
visa färre...
Abstract
Ämnesord
Stäng
- Surface Electromyography (sEMG) has applications in prosthetics, diagnostics and neuromuscular rehabilitation. Self-adhesive Ag/AgCl are the electrodes preferentially used to capture sEMG in short-term studies, however their long-term application is limited. In this study we designed and evaluated a fully integrated smart textile band with electrical connecting tracks knitted with intarsia techniques and knitted textile electrodes. Real-time myoelectric pattern recognition for motor volition and signal-to-noise ratio (SNR) were used to compare its sensing performance versus the conventional Ag-AgCl electrodes. After a comprehending measurement and performance comparison of the sEMG recordings, no significant differences were found between the textile and the Ag-AgCl electrodes in SNR and prediction accuracy obtained from pattern recognition classifiers.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Annan medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Other Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
Nyckelord
- Electrodes
- Real-time systems
- Pattern recognition
- Signal to noise ratio
- Electromyography
- Fabrics
- Applied Medical Technology
- Tillämpad medicinsk teknik