Scientific  instrument for a controlled choice of optimal photon energy in intra-oral radiography

Helmrot, Ebba (författare)
Östergötlands Läns Landsting,Radiofysikavdelningen,Landstinget i Östergötland
Sandborg, Michael, 1961- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Radiofysikavdelningen,Hälsouniversitetet
Eckerdal, Olle (författare)
n/a
visa fler...
Alm Carlsson, Gudrun (författare)
Linköpings universitet,Medicinsk radiofysik,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Hälsouniversitetet
visa färre...
 (creator_code:org_t)
1998
1998
Engelska.
Ingår i: Radiation Protection Dosimetry. - 0144-8420 .- 1742-3406. ; 80:1, s. 321-325
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Basic performance parameters are defined and analysed in order to optimise physical image quality in relation to the energy imparted to the patient in dental radiology. Air cavities were embedded in well-defined multimaterial, hard tissue phantoms to represent various objects in dento-maxillo-facial examinations. Basic performance parameters were: object contrast (C), energy imparted (_) to the patient, signal-to-noise ration (SNR), C2/_ (film) and (SNR)2/_ (digital imaging system) as functions of HVL (half-value layer), used to describe the photon energy spectrum. For the film receptor, the performance index C2/_ is maximum (optimal) at HVL values of 1.5-1.7 mm Al in the simulated Incisive, Premolar and Molar examinations. Other imaging tasks (examinations), not simulated here, may require other optimal HVL. For the digital imaging system (Digora) the performance index (SNR)2/_, theoretically calculated, indicates that a lower value of HVL is optimal than with film as receptor. However, due to the limited number of bits (8 bits) in the analogue to digital converter (ADC) contrast resolution is degraded and calls for use of higher photon energies (HVL). Customised optimisations with proper concern for patient category, type of examination, diagnostic task is the ultimate goal of this work. The conclusions stated above give some general advice on the appropriate choice of photon energy spectrum (HVL). In particular situations, it may be necessary to use more dose demanding kV settings (lower HVL) in order to get sufficient image quality for the diagnostic task.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy