SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:uu-353002"
 

Sökning: onr:"swepub:oai:DiVA.org:uu-353002" > Do dual-mobility cu...

Do dual-mobility cups cemented into porous tantalum shells reduce the risk of dislocation after revision surgery? : A retrospective cohort study on 184 patients

Brüggemann, Anders (författare)
Uppsala universitet,Ortopedi
Mallmin, Hans (författare)
Uppsala universitet,Ortopedi
Hailer, Nils P (författare)
Uppsala universitet,Ortopedi
 (creator_code:org_t)
2018-02-05
2018
Engelska.
Ingår i: Acta Orthopaedica. - : TAYLOR & FRANCIS LTD. - 1745-3674 .- 1745-3682. ; 89:2, s. 156-162
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background and purpose Dual-mobility cups (DMCs) reduce the risk of dislocation and porous tantalum (TM) shells show favorable osseointegration after acetabular revision surgery, yet the combination of these implants has not been studied. We hypothesized that (1) cementing a DMC into a TM shell decreases the risk of dislocation; (2) DMCs cemented into TM shells are not at greater risk of re-revision; (3) liberation of tantalum ions is marginal after use of this combined technique.Patients and methods We investigated the outcome in 184 hips (184 patients) after acetabular revision surgery with TM shells, fitted either with DMCs (n = 69), or with standard polyethylene (PE) liners (n = 115). Chart follow-up was complete for all patients, and the occurrence of dislocations and re-revisions was recorded. 20 were deceased, 50 were unable to attend follow-up, leaving 114 for assessment of hip function after 4.9 (0.5-8.9) years, radiographs were obtained in 99, and tantalum concentrations in 84 patients.Results 1 patient with a DMC had a dislocation, whereas 14 patients with PE liners experienced at least 1 dislocation. 11 of 15 re-revisions in the PE group were necessitated by dislocations, whereas none of the 2 re-revisions in the DMC group was performed for this reason. Hence, dislocation-free survival after 4 years was 99% (95% CI 96-100) in the DMC group, whereas it was 88% (CI 82-94, p = 0.01) in the PE group. We found no radiographic signs of implant failure in any patient. Mean tantalum concentrations were 0.1 mu l/L (CI 0.05-0.2) in the DMC group and 0.1 mu g/L (CI 0.05-0.2) in the PE group.Interpretation Cementing DMCs into TM shells reduces the risk of dislocation after acetabular revision surgery without jeopardizing overall cup survival, and without enhancing tantalum release.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Brüggemann, Ande ...
Mallmin, Hans
Hailer, Nils P
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Ortopedi
Artiklar i publikationen
Acta Orthopaedic ...
Av lärosätet
Uppsala universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy