Sökning: onr:"swepub:oai:DiVA.org:uu-430546" >
Fibula Graft Cuttin...
Fibula Graft Cutting Devices : Are 3D-Printed Cutting Guides More Precise than a Universal, Reusable Osteotomy Jig?
-
Meyer, Simon (författare)
-
- Hirsch, Jan-Michaél (författare)
- Uppsala universitet,Käkkirurgi,Department of Research & Development Public Dental Services Folktandvården AB, SE-118 27 Stockholm, Sweden
-
Leiggener, Christoph S. (författare)
-
visa fler...
-
Msallem, Bilal (författare)
-
Sigron, Guido R. (författare)
-
Kunz, Christoph (författare)
-
Thieringer, Florian M. (författare)
-
visa färre...
-
(creator_code:org_t)
- 2020-12-20
- 2020
- Engelska.
-
Ingår i: Journal of Clinical Medicine. - : MDPI AG. - 2077-0383. ; 9:12
- Relaterad länk:
-
https://doi.org/10.3...
-
visa fler...
-
https://uu.diva-port... (primary) (Raw object)
-
https://www.mdpi.com...
-
https://urn.kb.se/re...
-
https://doi.org/10.3...
-
visa färre...
Abstract
Ämnesord
Stäng
- Individual cutting guides for the reconstruction of lower jaw defects with fibular grafts are often used. However, the application of these osteotomy tools is costly and time intensive. The aim of this study was to compare the precision of osteotomies using a 3D-printed guide with those using a universal, reusable, and more cost-efficient Multi-Use Cutting Jig (MUC-Jig). In this non-blinded experimental study, 10 cranio-maxillofacial surgeons performed four graft removals each in a randomized order using the same osteotomy angle, both proximally (sagittal cut) and distally (coronal cut), of a graft (45°, 30°, 15°, or 0°), first with the MUC-Jig then with the 3D-printed cutting guide. The 40 fibula transplants (Tx) of each method (n = 80) were then analyzed concerning their Tx length and osteotomy angles and compared to the original planning data. Furthermore, the surgeons' subjective perception and the duration of the two procedures were analyzed. The mean relative length and mean relative angle deviation between the MUC-Jig (-0.08 ± 1.12 mm; -0.69° ± 3.15°) and the template (0.22 ± 0.90 mm; 0.36° ± 2.56°) group differed significantly (p = 0.002; p = < 0.001), but the absolute deviations did not (p = 0.206; p = 0.980). Consequently, clinically comparable osteotomy results can be achieved with both methods, but from an economic point of view the MUC-Jig is a more cost-efficient solution.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Odontologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dentistry (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kirurgi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Surgery (hsv//eng)
Nyckelord
- cranio-maxillofacial surgery
- fibula
- mandibular reconstruction
- osteotomy
- surgical guides
- tissue transplantation
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas