On the bone response to different implant textures. A 3D analysis of roughness, wavelength and surface pattern of experimental implants

Hallgren Höstner, Carin, 1971 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för biomaterialvetenskap,Institute of Surgical Sciences, Department of Biomaterials
 (creator_code:org_t)
ISBN 9162850113
Göteborg, 2001
Engelska.
  • Doktorsavhandling (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • PurposeThe purpose of the present thesis was to investigate the bone response to different isotropic, anisotropic and organised surface patterns.Material and methodsImplant surfaces have been modified with different techniques (abrasive blasting, turning, milling, photolithography or laser ablation). Turned implants were used as controls in all of the studies. The surface topographies have been characterised visually and numerically using instruments appropriate to the study (Mechanical Stylus Profilometer, Confocal Laser Scanning Profilometer and/or Atomic Force Microscopy).Five studies were undertaken using two implant designs; cylinders and screws. In studies I, IV and V a total of 160 commercially available threaded implants were used. In study II, 54 microimplants were used, and in study III 40 cylinders. All implants were produced from commercially pure (c. p.) titanium grade I or II. In total 27 human patients and 50 New Zealand White rabbits have been involved.After a healing period of 12 weeks the rabbits were sacrificed and the implants were evaluated with resonance frequency analysis (2 studies), peak removal torque (3 studies), pull-out (1 study) and histomorphometry. The human implants were evaluated with histomorphometry after a mean healing time of 6.3 months in the maxillae and 3.9 months in the mandibles.ResultsThe results indicate that the influence of waviness in the surface topography of implants is of less importance than the surface roughness. It also seems that an anisotropic surface is superior to an isotropic surface for bone response to an implant surface.ConclusionA blasted surface, which has a surface with no dominant pattern, seems to be preferable to an anisotropic surface or a surface with a chequered pattern.Increasing the average wavelength from 10 µm to approximately 40 µm did not improve implant incorporation in bone.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomaterialvetenskap (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomaterials Science (hsv//eng)

Nyckelord

Surface modification
surface texture
roughness
surface evaluation
in vivo studies.

Publikations- och innehållstyp

vet (ämneskategori)
dok (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Hallgren Höstner ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Medicinsk biotek ...
och Biomaterialveten ...
Av lärosätet
Göteborgs universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy