A two-compartment gel phantom for optimization and quality assurance in clinical BOLD fMRI

Olsrud, Johan (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Nilsson, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Mannfolk, Peter (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Waites, Anthony (författare)
Ståhlberg, Freddy (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
visa färre...
 (creator_code:org_t)
Elsevier BV, 2008
2008
Engelska.
Ingår i: Magnetic Resonance Imaging. - : Elsevier BV. - 1873-5894 .- 0730-725X. ; 26:2, s. 279-286
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Clinical applications of blood oxygen level-dependent (BOLD) functional magnetic resonance imaging (fMRI) depend heavily on robust paradigms, imaging methods and analysis procedures. In this work, as a means to optimize and perform quality assurance of the entire imaging and analysis chain, a phantom that provides a well known and reproducible signal change similar to a block type fMRI experiment is presented. It consists of two gel compartments with slightly different T(2) that dynamically enter and leave the imaged volume. The homogeneous gel in combination with a cylindrical geometry results in a well-defined T(2) difference causing a signal difference between the two compartments in T(2)-weighted MR images. From time series data obtained with the phantom, maps of percent signal change (PSC) and t-values are calculated. As an example of image parameter optimisation, the phantom is demonstrated to be useful for accurate determination of the influence of echo time (TE) on BOLD fMRI results, taking the t-value as a measure of sensitivity. In addition, the phantom is proposed as a tool for quality assurance (QA) since reproducible time series and t-maps are obtained in a series of independent repeat experiments. The phantom is relatively simple to build and can therefore be used by any clinical fMRI center.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Phantom
Agarose gel
BOLD
Optimization
fMRI

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy