SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:prod.swepub.kib.ki.se:144476319"
 

Sökning: onr:"swepub:oai:prod.swepub.kib.ki.se:144476319" > Augmented reality n...

Augmented reality navigation with intraoperative 3D imaging vs fluoroscopy-assisted free-hand surgery for spine fixation surgery: a matched-control study comparing accuracy

Elmi-Terander, A (författare)
Karolinska Institutet
Burstrom, G (författare)
Karolinska Institutet
Nachabe, R (författare)
visa fler...
Fagerlund, M (författare)
Stahl, F (författare)
Karolinska Institutet
Charalampidis, A (författare)
Karolinska Institutet
Edstrom, E (författare)
Karolinska Institutet
Gerdhem, P (författare)
Karolinska Institutet
visa färre...
 (creator_code:org_t)
2020-01-20
2020
Engelska.
Ingår i: Scientific reports. - : Springer Science and Business Media LLC. - 2045-2322. ; 10:1, s. 707-
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • This study aimed to compare screw placement accuracy and clinical aspects between Augmented Reality Surgical Navigation (ARSN) and free-hand (FH) technique. Twenty patients underwent spine surgery with screw placement using ARSN and were matched retrospectively to a cohort of 20 FH technique cases for comparison. All ARSN and FH cases were performed by the same surgeon. Matching was based on clinical diagnosis and similar proportions of screws placed in the thoracic and lumbosacral vertebrae in both groups. Accuracy of screw placement was assessed on postoperative scans according to the Gertzbein scale and grades 0 and 1 were considered accurate. Procedure time, blood loss and length of hospital stay, were collected as secondary endpoints. A total of 262 and 288 screws were assessed in the ARSN and FH groups, respectively. The share of clinically accurate screws was significantly higher in the ARSN vs FH group (93.9% vs 89.6%, p < 0.05). The proportion of screws placed without a cortical breach was twice as high in the ARSN group compared to the FH group (63.4% vs 30.6%, p < 0.0001). No statistical difference was observed for the secondary endpoints between both groups. This matched-control study demonstrated that ARSN provided higher screw placement accuracy compared to free-hand.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy