Robustness and lifetime of the bone conduction implant - a pilot study
-
- Fredén Jansson, Karl-Johan, 1988 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Håkansson, Bo, 1953 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
- Rigato, Cristina, 1989 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa fler...
-
- Eeg-Olofsson, Måns, 1967 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper,Institute of Clinical Sciences
-
- Reinfeldt, Sabine, 1978 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology
-
visa färre...
-
(creator_code:org_t)
- 2019
- 2019
- Engelska.
-
Ingår i: Medical Devices: Evidence and Research. - 1179-1470. ; 12, s. 89-100
- Relaterad länk:
-
https://research.cha... (primary) (free)
-
visa fler...
-
https://doi.org/10.2...
-
https://research.cha...
-
https://gup.ub.gu.se...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objectives: The objective of this study was to develop methods for evaluating the mechanical robustness and estimating the lifetime of the novel bone conduction implant (BCI) that is used in a clinical study. The methods are intended to be applicable to any similar device. Materials and methods: The robustness was evaluated using tests originally developed for cochlear implants comprising a random vibration test, a shock test, a pendulum test, and an impact test. Furthermore, magnetically induced torque and demagnetization during magnetic resonance imaging at 1.5 T were investigated using a dipole electromagnet. To estimate the lifetime of the implant, a long-term age-accelerated test was performed. Results: Out of all the tests, the pendulum and the impact tests had the largest effect on the electro-acoustic performance of the BCI implant, even if the change in performance was within acceptable limits (< 20%). In comparison with baseline data, the lower and higher resonance peaks shifted down in frequency by 13% and 18%, respectively, and with a loss in magnitude of 1.1 and 2.0 dB, respectively, in these tests. Conclusion: A complete series of tests were developed, and the BCI passed all the tests; its lifetime was estimated to be at least 26 years for patients who are using the implant for 12 hours on a daily basis.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Dermatologi och venereologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dermatology and Venereal Diseases (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Samhällsbyggnadsteknik -- Geoteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Civil Engineering -- Geotechnical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Oto-rhino-laryngologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Otorhinolaryngology (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Annan medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Other Medical Engineering (hsv//eng)
Nyckelord
- electro-acoustics
- electromagnetic transducer
- audiology
- bone conduction audiometry