Sökning: onr:"swepub:oai:research.chalmers.se:f17ea472-1506-48d7-b0b7-110f27f59bd3" >
3-D Microwave Tomog...
3-D Microwave Tomography Using the Soft Prior Regularization Technique: Evaluation in Anatomically Realistic MRI-Derived Numerical Breast Phantoms
-
- Golnabi, Amir H. (författare)
- Montclair State University
-
- Meaney, Paul M, 1960 (författare)
- Dartmouth College
-
- Geimer, Shireen D. (författare)
- Dartmouth College
-
visa fler...
-
- Paulsen, Keith D. (författare)
- Dartmouth College
-
visa färre...
-
(creator_code:org_t)
- 2019
- 2019
- Engelska.
-
Ingår i: IEEE Transactions on Biomedical Engineering. - 0018-9294 .- 1558-2531. ; 66:9, s. 2566-2575
- Relaterad länk:
-
https://research.cha...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objective: Fusion of magnetic resonance imaging (MRI) breast images with microwave tomography is accomplished through a soft prior technique, which incorporates spatial information (from MRI), i. e., accurate boundary location of different regions of interest, into the regularization process of the microwave image reconstruction algorithm. Methods: Numerical experiments were completed on a set of three-dimensional (3-D) breast geometries derived from MR breast data with different parenchymal densities, as well as a simulated tumor to evaluate the performance over a range of breast shapes, sizes, and property distributions. Results: When the soft prior regularization technique was applied, both permittivity and conductivity relative root mean square error values decreased by more than 87% across all breast densities, except in two cases where the error decrease was only 55% and 78%. In addition, the incorporation of structural priors increased contrast between tumor and fibroglandular tissue by 59% in permittivity and 192% in conductivity. Conclusion: This study confirmed that the soft prior algorithm is robust in 3-D and can function successfully across a range of complex geometries and tissue property distributions. Significance: This study demonstrates that our microwave tomography is capable of recovering accurate tissue property distributions when spatial information from MRI is incorporated through soft prior regularization.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
- NATURVETENSKAP -- Data- och informationsvetenskap -- Datorseende och robotik (hsv//swe)
- NATURAL SCIENCES -- Computer and Information Sciences -- Computer Vision and Robotics (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Image Processing (hsv//eng)
Nyckelord
- Breast cancer
- multimodality imaging
- soft prior regularization
- MRI
- microwave tomography
Publikations- och innehållstyp