Detection of mouth alcohol during breath alcohol analysis.
- 2015
-
Ingår i: Forensic Science International. - : Elsevier BV. - 1872-6283 .- 0379-0738. ; 249, s. 66-72
-
Tidskriftsartikel (refereegranskat)abstract
- The presence of mouth alcohol (MA) during alcohol breath test for law enforcement is the most common cause of falsely high breath alcohol concentrations (BrAC). A fast and reliable test for detection of MA roadside at the scene of the act would facilitate the police efforts for proper prosecution. A tentative technique to use orally exhaled water vapour as a reference gas to position the origin of alcohol was validated. BrAC and water vapour concentration (WVC) were simultaneously measured as a known MA component was added to subjects with existing blood alcohol. In the absence of MA, water always precedes alcohol in a volumetric expirogram. In the presence of MA this relationship reversed. A scatterplot of WVC versus BrAC from similar fractional exhaled volumes illustrates how their relative positions change by MA. A deviation area (DA) between the scatterplot curve and a fictitious linear relationship was defined as a measurement of MA. The accuracy and cut-off level of the DA to detect MA were determined with receiver operating characteristic (ROC) curve analysis. The area under the ROC curve (AUC) was 0.95 (95% CI 0.90-1.0), indicating excellent discriminatory ability. The optimal cut-off for DA to discriminate between MA ≥0.010mg/L (1μg/100ml, 0.002g/210L) or lack of MA was -0.35, with a sensitivity of 0.91 and specificity of 0.95. Analysis of BrAC in relation to WVC is a practical method to detect and confirm MA contamination with high reliability.
|
|