1. |
- Khampha, Wanida, et al.
(författare)
-
Specific detection of L-glutamate in food using flow-injection analysis and enzymatic recycling of substrate
- 2004
-
Ingår i: Analytica Chimica Acta. - : Elsevier BV. - 1873-4324 .- 0003-2670. ; 518:1-2, s. 127-135
-
Tidskriftsartikel (refereegranskat)abstract
- A flow injection analysis (FIA) system for specific determination of L-glutamate in food samples based on a bi-enzymatic amplification system has been developed. The content of L-glutamate in the sample was amplified by cycling between L-glutamate dehydrogenase (GIDH) and a novel enzyme, D-phenylglycine aminotransferase (D-PhgAT). In this system, GIDH converts L-glutamate to 2-oxoglutarate with concomitant reduction of NAD(+) to NADH. D-PhgAT transfers an amino group from D-4-hydroxyphenylglycine to 2-oxoglutarate, thus recycling L-glutamate. Accumulation of NADH in the course of the enzymatic recycling was monitored both by fluorescence and UV absorbance and used for quantification of L-glutamate. The assay was characterized by high long-term stability (at least 70 days) and good reproducibility (within-day and between-day RSDs were 4.3-7.3% and 8.9%). The fluorimetric assay was slightly more sensitive with a L-glutamate detection limit of 0.4 muM and linear range of 2.5-50 muM. The assay was specific for L-glutamate, with recoveries between 95-103% in the presence of 17 different amino acids tested one by one. The method was applied to analysis of real food samples and results were correlated with a commercial Boehringer Mannheim assay kit. (C) 2004 Elsevier B.V. All rights reserved.
|
|