In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty

Goldvasser, Dov (author)
Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston MA, United States
Hansen, Viktor J. (author)
Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston MA, United States
Noz, Marilyn E. (author)
New York University, Department of Radiology,Nuclear Medicine
show more...
Maguire Jr., Gerald Q., Jr (author)
KTH,Radio Systems Laboratory (RS Lab)
Zeleznik, Michael P. (author)
University of Utah, Salt Lake City, UT, USA,School of Computing
Olivecrona, Henrik (author)
Karolinska Institute,Department of Molecular Medicine and Surgery, Section of Orthopaedics and Sport Medicine
Bragdon, Charles R. (author)
Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston MA, United States
Weidenhielm, Lars (author)
Karolinska Institute,Department of Molecular Medicine and Surgery
Malchau, Henrik (author)
Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston MA, United States
show less...
 (creator_code:org_t)
2014-04-23
2014
English.
In: Acta Orthopaedica. - : Medical Journals Sweden AB. - 1745-3674 .- 1745-3682. ; 85:3, s. 271-275
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Background - Determination of the amount of wear in a polyethylene liner following total hip arthroplasty (THA) is important for both the clinical care of individual patients and the development of new types of liners. Patients and methods - We measured in vivo wear of the polyethylene liner using computed tomography (CT) (obtained in the course of regular clinical care) and compared it to coordinate-measuring machine (CMM) readings. Also, changes in liner thickness of the same retrieved polyethylene liner were measured using a micrometer, and were compared to CT and CMM measurements. The distance between the centers of the acetabular cup and femoral head component was measured in 3D CT, using a semi-automatic analysis method. CMM readings were performed on each acetabular liner and data were analyzed using 3D computer-aided design software. Micrometer readings compared the thickest and thinnest regions of the liner. We analyzed 10 THA CTs and retrievals that met minimal requirements for CT slice thickness and explanted cup condition. Results - For the 10 cups, the mean difference between the CT readings and the CMM readings was -0.09 (-0.38 to 0.20) mm. This difference was not statistically significant (p = 0.6). Between CT and micrometer, the mean difference was 0.11 (-0.33 to 0.55) mm. This difference was not statistically significant (p = 0.6). Interpretation - Our results show that CT imaging is ready to be used as a tool in clinical wear measurement of polyethylene liners used in THA.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Keyword

Cross-Linked Polyethylene
Computed-Tomography
CT
Trial

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view