X-ray phase contrast for CO2 microangiography
-
- Lundström, Ulf (författare)
- KTH,Biomedicinsk fysik och röntgenfysik
-
- Larsson, Daniel H. (författare)
- KTH,Biomedicinsk fysik och röntgenfysik
-
- Burvall, Anna (författare)
- KTH,Biomedicinsk fysik och röntgenfysik
-
visa fler...
-
- Takman, Per A. C. (författare)
- KTH,Biomedicinsk fysik och röntgenfysik
-
- Scott, L. (författare)
- Karolinska Institutet
-
- Brismar, H. (författare)
- Karolinska Institutet
-
- Hertz, Hans M. (författare)
- KTH,Biomedicinsk fysik och röntgenfysik
-
visa färre...
-
(creator_code:org_t)
- 2012-04-13
- 2012
- Engelska.
-
Ingår i: Physics in Medicine and Biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 57:9, s. 2603-2617
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- We demonstrate a laboratory method for imaging small blood vessels using x-ray propagation-based phase-contrast imaging and carbon dioxide (CO2) gas as a contrast agent. The limited radiation dose in combination with CO2 being clinically acceptable makes the method promising for small-diameter vascular visualization. We investigate the possibilities and limitations of the method for small-animal angiography and compare it with conventional absorption-based x-ray angiography. Photon noise in absorption-contrast imaging prevents visualization of blood vessels narrower than 50 mu m at the highest radiation doses compatible with living animals, whereas our simulations and experiments indicate the possibility of visualizing 20 mu m vessels at radiation doses as low as 100 mGy. Experimental computed tomography of excised rat kidney shows blood vessels of diameters down to 60 mu m with improved image quality compared to absorption-based methods. With our present prototype x-ray source, the acquisition time for a tomographic dataset is approximately 1 h, which is long compared to the 1-20 min common for absorption-contrast micro-CT systems. Further development of the liquid-metal-jet microfocus x-ray sources used here and high-resolution x-ray detectors shows promise to reduce exposure times and make this high-resolution method practical for imaging of living animals.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Angiography
- Blood vessels
- Carbon dioxide
- Dosimetry
- Image quality
- Imaging techniques
- Visualization
- X ray apparatus
- X rays