SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:oru-56419"
 

Sökning: onr:"swepub:oai:DiVA.org:oru-56419" >

A visual grading study for different administered activity levels in bone scintigraphy

Gustafsson, Agneta (författare)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US
Karlsson, Henrik (författare)
Department of Radiation Physics, Kalmar County Hospital, Kalmar, Sweden
Olsson, Anna (författare)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
visa fler...
Geijer, Håkan, 1961- (författare)
Region Örebro län,Department of Radiology,Örebro University Hospital, Sweden
Olsson, Anna (författare)
Department of Medical Physics, IMH, Linköping University, Linköping, Sweden; Radiation Physics UHL, County council of Östergötland, Linköping, Sweden; Center for Medical Image Science and Visualization, Linköping University, Linköping, Sweden
visa färre...
 (creator_code:org_t)
2014-05-04
Engelska.
Ingår i: Clinical Physiology and Functional Imaging. - : Wiley-Blackwell. - 1475-0961 .- 1475-097X. ; 35:3, s. 231-236
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Introduction: The aim of the study is to assess the administered activity levels versus visual-based image quality using visual grading regression (VGR) including an assessment of the newly stated image criteria for whole-body bone scintigraphy.Materials and methods: A total of 90 patients was included and grouped in three levels of administered activity: 400, 500 and 600 MBq. Six clinical image criteria regarding image quality was formulated by experienced nuclear medicine physicians. Visual grading was performed in all images, where three physicians rated the fulfilment of the image criteria on a four-step ordinal scale. The results were analysed using VGR. A count analysis was also made where the total number of counts in both views was registered.Results: The administered activity of 600 MBq gives significantly better image quality than 400 MBq in five of six criteria (P<005). Comparing the administered activity of 600 MBq to 500 MBq, four criteria of six show significantly better image quality (P<005). The administered activity of 500 MBq gives no significantly better image quality than 400 Mbq (P<005). The count analysis shows that none of the three levels of administrated activity fulfil the recommendations by the EANM.Conclusion: There was a significant improvement in perceived image quality using an activity level of 600 MBq compared to lower activity levels in whole-body bone scintigraphy for the gamma camera equipment end set-up used in this study. This type of visual-based grading study seems to be a valuable tool and easy to implement in the clinical environment.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinska och farmaceutiska grundvetenskaper -- Fysiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Basic Medicine -- Physiology (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Nyckelord

bone and bones
evaluation
radionuclide imaging
technetium Tc 99 m

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Gustafsson, Agne ...
Karlsson, Henrik
Olsson, Anna
Geijer, Håkan, 1 ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Medicinska och f ...
och Fysiologi
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
Artiklar i publikationen
Clinical Physiol ...
Av lärosätet
Örebro universitet
Linköpings universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy