Influence of anode-filter combinations on image quality and radiation dose in 965 women undergoing mammography
-
- Thilander-Klang, Anne, 1959 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
-
Ackerholm, Per (författare)
-
- Berlin, Isabelita (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiologi,Institute of Selected Clinical Sciences, Department of Radiology
-
visa fler...
-
- Bjurstam, Nils (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiologi,Institute of Selected Clinical Sciences, Department of Radiology
-
Mattsson, Sören (författare)
-
- Månsson, Lars Gunnar, 1951 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för särskilda specialiteter, Avdelningen för radiofysik,Institute of Selected Clinical Sciences, Department of Radiation Physics
-
von Schéele, Christina (författare)
-
Thunberg, Stefan (författare)
-
visa färre...
-
(creator_code:org_t)
- 1997
- 1997
- Engelska.
-
Ingår i: Radiology. - 0033-8419. ; 203, s. 348-354
- Relaterad länk:
-
https://gup.ub.gu.se...
Abstract
Ämnesord
Stäng
- PURPOSE: To evaluate how anode-filter combinations influence image quality in and mean glandular dose to breasts of different thicknesses and compositions. MATERIALS AND METHODS: Mammograms were obtained with a molybdenum (Mo) anode and a Mo filter at 26 kVp, a Mo anode and a rhodium (Rh) filter at 27 kVp, or a tungsten (W) anode and a Rh filter at 26 kVp in 965 women. One anode-filter-tube voltage combination was used in the right breast and another in the left. The mean glandular dose to each breast was calculated. RESULTS: Image contrast was highest in the Mo-Mo mammograms. However, depiction of the glandular tissue, pectoral muscle, and skin and subcutis was significantly (P < .001) better with the Mo-Rh and the W-Rh than with the Mo-Mo combination. The average mean absorbed doses to the glandular tissue for W-Rh and Mo-Rh were 50% and 75%, respectivety, of that for Mo-Mo. CONCLUSION: Breast thickness is the most important parameter in setection of an anode-filter-tube voltage combination. Compared with Mo-Mo, both Mo-Rh and W-Rh gave good image quality of the mammary gland and a considerably lower absorbed dose. Mo-Rh-27 kVp is recommended for breast thicknesses of 60 mm or less; W-Rh-26 kVp, for breast thicknesses of greater than 60 mm.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Breast radiography
- quality assurance
- 00.112 Breast radiography
- radiation dose
- 00.112 Images
- quality
- 00.112