Aneurysm clip MR artifacts. Titanium versus stainless steel and influence of imaging parameters

Björkman-Burtscher, Isabella (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Owman, Titti (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Romner, Bertil (författare)
Lund University,Lunds universitet,Neurokirurgi,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Neurosurgery,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Ståhlberg, Freddy (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Holtås, Stig (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa färre...
 (creator_code:org_t)
1998
1998
Engelska.
Ingår i: Acta Radiologica. - 1600-0455. ; 39:1, s. 70-76
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: The aim of this study was to evaluate the extent to which titanium aneurysm clips could improve the quality of MR imaging compared with stainless steel clips, and to determine whether the clip artifacts could be reduced by controlling certain MR imaging parameters in frequently used pulse sequences. MATERIAL AND METHODS: The metal artifacts induced by 3 aneurysm clips were compared in 3 pulse sequences. The clips were: a Yasargil titanium aneurysm clip FT 752 T; a Yasargil standard aneurysm clip FE 752 K; and, for comparison, a ferromagnetic Scoville aneurysm clip En-58J. The pulse sequences were: spin echo (SE); gradient echo (GE); and fast SE. An evaluation was made of 3 imaging parameters with regard to their influence on the size of the metal artifacts. The parameters were: bandwidth; echo time (TE); and echo-train length. RESULTS: The titanium clip showed artifacts that were about 60% smaller than those from the stainless steel clip. The only parameter that influenced artifact size to any major degree was bandwidth in the SE sequences but not in the GE sequences. GE sequences induced larger artifacts than SE sequences and showed larger artifacts with longer TE. CONCLUSION: Titanium aneurysm clips reduced MR artifacts by approximately 60% compared to stainless steel clips. Artifacts were further reduced by using SE-based sequences with a high bandwidth or, if necessary, GE sequences with a low TE.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy