Repeatability and reproducibility of longitudinal relaxation rate in 12 small-animal MRI systems

Waterton, John C. (författare)
University of Manchester,Bioxydyn Ltd
Hines, Catherine D.G. (författare)
Merck Sharp And Dohme Corp., US
Hockings, Paul, 1956 (författare)
Chalmers University of Technology,Antaros Medical AB
visa fler...
Laitinen, Iina (författare)
Sanofi-Aventis Deutschland GmbH
Ziemian, Sabina (författare)
Bayer AG,Bayer Pharma AG
Campbell, Simon (författare)
GlaxoSmithKline,SmithKline Beecham Pharmaceuticals
Gottschalk, Michael (författare)
Lund University,Lunds universitet,Lund University Bioimaging Center,Medicinska fakulteten,MR Physics,Forskargrupper vid Lunds universitet,Faculty of Medicine,Lund University Research Groups
Green, Claudia (författare)
Bayer AG,Bayer Pharma AG
Haase, Michael (författare)
GlaxoSmithKline,SmithKline Beecham Pharmaceuticals
Hassemer, Katja (författare)
Sanofi-Aventis Deutschland GmbH
Juretschke, Hans Paul (författare)
Sanofi-Aventis Deutschland GmbH
Koehler, Sascha (författare)
Bruker BioSpin GmbH, Germany,Bruker BioSpin MRI GmbH
Lloyd, William (författare)
University of Manchester
Luo, Yanping (författare)
AbbVie,AbbVie Inc.
Mahmutovic Persson, Irma (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
O'Connor, James P.B. (författare)
University of Manchester
Olsson, Lars E. (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Pindoria, Kashmira (författare)
GlaxoSmithKline,SmithKline Beecham Pharmaceuticals
Schneider, Jurgen E. (författare)
University of Leeds
Sourbron, Steven (författare)
University of Leeds
Steinmann, Denise (författare)
Sanofi-Aventis Deutschland GmbH
Strobel, Klaus (författare)
Bruker BioSpin GmbH, Germany,Bruker BioSpin MRI GmbH
Tadimalla, Sirisha (författare)
University of Leeds
Teh, Irvin (författare)
University of Leeds
Veltien, Andor (författare)
Radboud Universiteit,Radboud University,Radboud University Medical Center
Zhang, Xiaomeng (författare)
AbbVie,AbbVie Inc.
Schütz, Gunnar (författare)
Bayer AG,Bayer Pharma AG
visa färre...
 (creator_code:org_t)
Elsevier BV, 2019
2019
Engelska.
Ingår i: Magnetic Resonance Imaging. - : Elsevier BV. - 1873-5894 .- 0730-725X. ; 59, s. 121-129
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Many translational MR biomarkers derive from measurements of the water proton longitudinal relaxation rate R 1 , but evidence for between-site reproducibility of R 1 in small-animal MRI is lacking. Objective: To assess R 1 repeatability and multi-site reproducibility in phantoms for preclinical MRI. Methods: R 1 was measured by saturation recovery in 2% agarose phantoms with five nickel chloride concentrations in 12 magnets at 5 field strengths in 11 centres on two different occasions within 1–13 days. R 1 was analysed in three different regions of interest, giving 360 measurements in total. Root-mean-square repeatability and reproducibility coefficients of variation (CoV) were calculated. Propagation of reproducibility errors into 21 translational MR measurements and biomarkers was estimated. Relaxivities were calculated. Dynamic signal stability was also measured. Results: CoV for day-to-day repeatability (N = 180 regions of interest) was 2.34% and for between-centre reproducibility (N = 9 centres) was 1.43%. Mostly, these do not propagate to biologically significant between-centre error, although a few R 1 -based MR biomarkers were found to be quite sensitive even to such small errors in R 1 , notably in myocardial fibrosis, in white matter, and in oxygen-enhanced MRI. The relaxivity of aqueous Ni 2+ in 2% agarose varied between 0.66 s −1 mM −1 at 3 T and 0.94 s −1 mM −1 at 11.7T. Interpretation: While several factors affect the reproducibility of R 1 -based MR biomarkers measured preclinically, between-centre propagation of errors arising from intrinsic equipment irreproducibility should in most cases be small. However, in a few specific cases exceptional efforts might be required to ensure R 1 -reproducibility.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Phantom
Relaxation time
Reproducibility
Biomarker
MRI
Hardware stability
Error propagation
Biomarker
Error propagation
Hardware stability
MRI
Phantom
Relaxation time
Reproducibility

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy