Final report from NMR project 'Test of a filter pack combined with a PM10 inlet'. Project number 04FOX5

Ferm, Martin (författare)
IVL Svenska Miljöinstitutet
Makkonen, Ulla (författare)
Hanssen, Jan Erik (författare)
visa fler...
Ellermann, Thomas (författare)
visa färre...
 (creator_code:org_t)
IVL Svenska Miljöinstitutet, 2006
Engelska.
Serie: B-rapport ; B1686
  • Rapport (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • The EMEP filter pack for measuring gaseous and particulate reduced and oxidised nitrogen compounds in air has been modified with a PM10 inlet to meet the EU air quality directive for particle measurements. The filter pack also collects sulphur compounds as well as base cations and sodium chloride. The EMEP filter pack has been used for 20 years now and it is important that the PM10 inlet doesn't affect the results. The cut-off for the filter pack recommended by EMEP is unknown and may be affected by wind. Here it was found that the cut-off for the EMEP filter pack is larger than it is for PM10. This implies that higher calcium concentrations will be obtained with the original EMEP filter pack without a PM10 inlet. All comparisons of the two filter packs show excellent agreement for sulphur dioxide and man made sulphate particles. Some losses of total nitrate have been found with the PM10 inlet especially of particulate nitrate. Some losses of particulate ammonium also take place but do not affect the total concentration.
  • The EMEP filter pack for measuring gaseous and particulate reduced and oxidised nitrogen compounds in air has been modified with a PM10 inlet to meet the EU air quality directive for particle measurements. The filter pack also collects sulphur compounds as well as base cations and sodium chloride. The EMEP filter pack has been used for 20 years now and it is important that the PM10 inlet doesn't affect the results. The cut-off for the filter pack recommended by EMEP is unknown and may be affected by wind. Here it was found that the cut-off for the EMEP filter pack is larger than it is for PM10. This implies that higher calcium concentrations will be obtained with the original EMEP filter pack without a PM10 inlet. All comparisons of the two filter packs show excellent agreement for sulphur dioxide and man made sulphate particles. Some losses of total nitrate have been found with the PM10 inlet especially of particulate nitrate. Some losses of particulate ammonium also take place but do not affect the total concentration.

Nyckelord

EMEP
PM10
filter pack
nitric acid
nitrate
ammonia
ammonium
base cations

Publikations- och innehållstyp

vet (ämneskategori)
rap (ämneskategori)

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Ferm, Martin
Makkonen, Ulla
Hanssen, Jan Eri ...
Ellermann, Thoma ...
Delar i serien
Av lärosätet
IVL Svenska Miljöinstitutet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy