Why ultra high performance liquid chromatography produces more tailing peaks than high performance liquid chromatography, why it does not matter and how it can be addressed

Petersson, Patrik (författare)
AstraZeneca R&D Lund, SE-221 87 Lund, Sweden
Forssén, Patrik (författare)
Karlstads universitet,Avdelningen för kemi och biomedicinsk vetenskap,Department of Chemistry and Biomedical Science, Karlstad University, Sweden
Edström, Lena (författare)
Uppsala universitet,Analytisk kemi,Department of Physical and Analytical Chemistry, BMC Box 577, SE-751 23 Uppsala, Sweden
visa fler...
Samie, Farzad (författare)
AstraZeneca Nordic Headquarters (ISMO), Södertälje,, SE-151 85 Sweden,AstraZeneca Nordic Headquarters, Södertälje, Sweden
Tatterton, Stephen (författare)
AstraZeneca R&D Charnwood, Loughborough, Leicestershire LE11 5RH, UK
Clarke, Adrian (författare)
AstraZeneca R&D Charnwood, Loughborough, Leicestershire LE11 5RH, UK
Fornstedt, Torgny (författare)
Karlstads universitet,Avdelningen för kemi och biomedicinsk vetenskap,Department of Chemistry and Biomedical Science, Karlstad University, Sweden
visa färre...
 (creator_code:org_t)
Elsevier BV, 2011
2011
Engelska.
Ingår i: Journal of chromatography. B. - : Elsevier BV. - 1570-0232 .- 1873-376X .- 0021-9673 .- 1873-3778. ; 1218:39, s. 6914-6921
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The purpose of this study is to demonstrate, with experiments and with computer simulations based on a firm chromatographic theory, that the wide spread perception of that the United States Pharmacopeia tailing factor must be lower than 2 (Tf < 2) is questionable when using the latest generation of LC equipment. It is shown that highly efficient LC separations like those obtained with sub-2 μm porous and 2.7 μm superficially porous particles (UHPLC) produce significantly higher Tf-values than the corresponding separation based on 3 μm porous particles (HPLC) when the same amount of sample is injected. Still UHPLC separations provide a better resolution to adjacent peaks. Expressions have been derived that describe how the Tf-value changes with particle size or number of theoretical plates. Expressions have also been derived that can be used to scale the injection volume based on particle size or number of theoretical plates to maintain the Tf-value when translating a HPLC separation to the corresponding UHPLC separation. An aspect that has been ignored in previous publications. Finally, data obtained from columns with different age/condition indicate that Tf-values should be complemented by a peak width measure to provide a more objective quality measure.

Ämnesord

NATURVETENSKAP  -- Kemi -- Analytisk kemi (hsv//swe)
NATURAL SCIENCES  -- Chemical Sciences -- Analytical Chemistry (hsv//eng)
NATURVETENSKAP  -- Kemi (hsv//swe)
NATURAL SCIENCES  -- Chemical Sciences (hsv//eng)

Nyckelord

HPLC
Isotherm
Particle size
Simulation
Translation
UHPLC
Chemistry

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy