Evaluation of ultrasound Tissue Velocity Imaging : a phantom study of velocity estimation in skeletal muscle low-level contractions

Lindberg, Frida, 1982- (författare)
KTH,Medicinsk teknik
Mårtensson, Mattias (författare)
KTH,Medicinsk bildteknik
Grönlund, Christer (författare)
Umeå universitet,Radiofysik,Medicinsk Teknik - FoU, Umeå Univeristet
visa fler...
Brodin, Lars-Åke (författare)
KTH,Medicinsk teknik
visa färre...
 (creator_code:org_t)
2013-06-07
2013
Engelska.
Ingår i: BMC Medical Imaging. - : BioMed Central. - 1471-2342. ; 13:1, s. 16-
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. Methods: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. Results: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. Conclusions: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Tissue Velocity Imaging
Ultrasound
Skeletal muscle
Phantom evaluation
Pulse repetition frequency

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy