Single-shot dual-energy subtraction mammography with electronic spectrum splitting : Feasibility

Bornefalk, Hans (författare)
Lewin, John M. (författare)
Danielsson, Mats (författare)
visa fler...
Lundqvist, Mats (författare)
visa färre...
Elsevier BV, 2006
2006
Engelska.
Ingår i: European Journal of Radiology. - : Elsevier BV. - 0720-048X .- 1872-7727. ; 60:2, s. 275-278
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We present a single-shot dual-energy subtraction mammography technique using an energy sensitive photon counting detector. An electronic threshold near the middle of the X-ray spectrum discriminates between high- and low-energy photons, and allows the simultaneous acquisition of high- and low-energy images which can be combined to suppress anatomical clutter. By setting the electronic threshold close to 33.2 keV (the k-edge of iodine) the system is optimized for dual-energy contrast-enhanced imaging of breast tumors. This method eliminates the need for separate exposures which might otherwise lead to motion artifacts. The method is illustrated in phantom images.

Nyckelord

dual-energy imaging
digital mammography
contrast-enhancement
physics
breast
breast-diagnosis
breast radiography
technology
digital mammography
contrast-media
experience

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy