Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups : An Experimental Study

Brodén, Cyrus (författare)
Department of Molecular Medicine and Surgery, Karolinska Institutet
Olivecrona, Henrik (författare)
Department of Molecular Medicine and Surgery, Karolinska Institutet
Maguire Jr., Gerald Q. (författare)
KTH,Radio Systems Laboratory (RS Lab)
visa fler...
Noz, Marilyn E. (författare)
New York University, Department of Radiology
Zeleznik, Michael P. (författare)
School of Computing, College of Engineering, University of Utah
Sköldenberg, Olof (författare)
Karolinska Institutet
visa färre...
 (creator_code:org_t)
Hindawi Publishing Corporation, 2016
2016
Engelska.
Ingår i: BioMed Research International. - : Hindawi Publishing Corporation. - 2314-6133 .- 2314-6141.
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background and Purpose. The gold standard for detection of implant wear and migration is currently radiostereometry (RSA). The purpose of this study is to compare a three-dimensional computed tomography technique (3D CT) to standard RSA as an alternative technique for measuring migration of acetabular cups in total hip arthroplasty.Materials and Methods. With tantalum beads, we marked one cemented and one uncemented cup and mounted these on a similarly marked pelvic model. A comparison was made between 3D CT and standard RSA for measuring migration. Twelve repeated stereoradiographs and CT scans with double examinations in each position and gradual migration of the implants were made. Precision and accuracy of the 3D CT were calculated.Results. The accuracy of the 3D CT ranged between 0.07 and 0.32 mm for translations and 0.21 and 0.82° for rotation. The precision ranged between 0.01 and 0.09 mm for translations and 0.06 and 0.29° for rotations, respectively. For standard RSA, the precision ranged between 0.04 and 0.09 mm for translations and 0.08 and 0.32° for rotations, respectively. There was no significant difference in precision between 3D CT and standard RSA. The effective radiation dose of the 3D CT method, comparable to RSA, was estimated to be 0.33 mSv.Interpretation. Low dose 3D CT is a comparable method to standard RSA in an experimental setting.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy