SwePub
Sök i LIBRIS databas

  Extended search

id:"swepub:oai:DiVA.org:kth-315905"
 

Search: id:"swepub:oai:DiVA.org:kth-315905" > Safety of arterial ...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist
  • Nordenfur, Tim,1990-KTH,Medicinteknik och hälsosystem (author)

Safety of arterial shear wave elastography-ex-vivo assessment of induced strain and strain rates

  • Article/chapterEnglish2022

Publisher, publication year, extent ...

  • 2022-07-19
  • IOP Publishing,2022
  • printrdacarrier

Numbers

  • LIBRIS-ID:oai:DiVA.org:kth-315905
  • https://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-315905URI
  • https://doi.org/10.1088/2057-1976/ac7f39DOI
  • http://kipublications.ki.se/Default.aspx?queryparsed=id:235797069URI

Supplementary language notes

  • Language:English
  • Summary in:English

Part of subdatabase

Classification

  • Subject category:ref swepub-contenttype
  • Subject category:art swepub-publicationtype

Notes

  • QC 20220728
  • Shear wave elastography (SWE) is a promising technique for characterizing carotid plaques and assessing local arterial stiffness. The mechanical stress to which the tissue is subjected during SWE using acoustic radiation force (ARF), leading to strain at a certain strain rate, is still relatively unknown. Because SWEis increasingly used for arterial applications where the mechanical stress could potentially lead to significant consequences, it is important to understand the risks of SWE-induced strain and strain rate. The aim of this study was to investigate the safety of SWE in terms of induced arterial strain and strain rate ex-vivo and in a human carotid artery in-vivo. SWE was performed on six porcine aortae as a model of the human carotid artery using different combinations of ARF push parameters (push voltage: 60/90 V, aperture width: f/1.0/1.5, push length: 100/150/200 mu s) and distance to push position. The largest induced strain and strain rate were 1.46% and 54 s(-1) (90 V, f/ 1.0, 200 mu s), respectively. Moreover, the SWE-induced strains and strain rates increased with increasing push voltage, aperture, push length, and decreasing distance between the region of interest and the push. In the human carotid artery, the SWE-induced maximum strain was 0.06% and the maximum strain rate was 1.58 s(-1), compared with the maximum absolute strain and strain rate of 12.61% and 5.12 s(-1), respectively, induced by blood pressure variations in the cardiac cycle. Our results indicate that ex-vivo arterial SWE does not expose the artery to higher strain rate than normal blood pressure variations, and to strain one order of magnitude higher than normal blood pressure variations, at the push settings and distances from the region of interest used in this study.

Subject headings and genre

Added entries (persons, corporate bodies, meetings, titles ...)

  • Caidahl, KennethKarolinska Institutet (author)
  • Grishenkov, Dmitry,Associate Professor,1983-KTH,Medicinsk avbildning(Swepub:kth)u1b9avjc (author)
  • Maksuti, EliraKarolinska Inst, Dept Physiol & Pharmacol, Anaesthesiol & Intens Care, Stockholm, Sweden. (author)
  • Marlevi, DavidKarolinska Institutet (author)
  • Urban, Matthew W.Mayo Clin, Dept Radiol, Rochester, MN USA. (author)
  • Larsson, Matilda,1981-KTH,Medicinsk avbildning,Dept Biomed Engn & Hlth Syst, Stockholm, Sweden(Swepub:kth)u1kv408t (author)
  • KTHMedicinteknik och hälsosystem (creator_code:org_t)

Related titles

  • In:Biomedical Engineering & Physics Express: IOP Publishing8:52057-1976

Internet link

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view