Comparison of heart rate measured by Polar RS400 and ECG, validity and repeatability

Engstrom, Elise (author)
Division of Physiotherapy, Department of Health Sciences, Lund University, Sweden
Ottosson, Elin (author)
Division of Physiotherapy, Department of Health Sciences, Lund University,Sweden
Wohlfart, Björn (author)
Division of Thoracic Surgery, Department of Clinical Sciences, Lund University and Igelösa Medical Clinic, Lund, Sweden
show more...
Grundstrom, Nils (author)
Linköpings universitet,Avdelningen för läkemedelsforskning,Hälsouniversitetet
Wisen, Anita (author)
Division of Physiotherapy, Department of Health Sciences, Lund University, Sweden
show less...
 (creator_code:org_t)
2012-06-25
2012
English.
In: Advances in Physiotherapy. - : Informa Healthcare. - 1403-8196 .- 1651-1948. ; 14:3, s. 115-122
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Aims: The purpose of this study was to investigate criterion-related validity and testretest repeatability of the heart rate monitor Polar RS400 versus electrocardiogram (ECG). Methodology: Ten healthy subjects, 19-34 years, performed a cycle ergometer test 5 min on each load (50, 100 and 150 W). Heart rate (HR) was measured with ECG and Polar RS400 and recorded digitally. After at least one hour resting the test was repeated. Major findings: The results showed a significant correlation between HR measured by ECG and by Polar RS400 with correlation coefficients ranging from 0.97 to 1.00. In test 1 the mean difference ±2SD between HR Polar and HR ECG was 0.7±4.3 bpm and in test 2, 0.2±3.2 bpm. In the repeated tests, the mean difference of HR between test 2 and test 1±2SD was 3.2±11.9 bpm with ECG and 2.6±14.3 bpm with Polar RS400 and these differences were not statistically significant. Conclusion: This study indicates good criterion-related validity and testretest repeatability of Polar RS400. Differences observed at individual levels should be noticed, but are not considered to be clinically important. Polar RS400 is thus well suited for recording HR during physical activity and exercise training. © 2012 Informa Healthcare.

Keyword

ergometer test; exercise; heart rate monitoring; physiotherapy; submaximal activity
MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view