Test-retest reliability of automated whole body and compartmental muscle volume measurements on a wide bore 3T MR system

Thomas, Marianna S (författare)
Department of Radiology, Norfolk & Norwich University Hospital, Norwich, UK.
Newman, David (författare)
Department of Radiology, Norfolk & Norwich University Hospital, Norwich, UK.
Dahlqvist Leinhard, Olof (författare)
Östergötlands Läns Landsting,Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Radiofysikavdelningen US
visa fler...
Karlsson, Anette (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Medicinsk informatik,Tekniska högskolan
Rosander, Johannes (författare)
Advanced MR Analytics AB, Linköping, Sweden
Kasmai, Bahman (författare)
Department of Radiology Norfolk & Norwich University Hospital, United Kingdom
Greenwood, Richard (författare)
Department of Radiology Norfolk & Norwich University Hospital, United Kingdom
Malcolm, Paul N (författare)
Department of Radiology Norfolk & Norwich University Hospital, United Kingdom
Borga, Magnus (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Medicinsk informatik,Tekniska högskolan
Toms, Andoni P (författare)
Department of Radiology Norfolk & Norwich University Hospital, United Kingdom
visa färre...
Department of Radiology, Norfolk & Norwich University Hospital, Norwich, UK Centrum för medicinsk bildvetenskap och visualisering, CMIV (creator_code:org_t)
2014-05-29
2014
Engelska.
Ingår i: European Radiology. - : Springer Berlin/Heidelberg. - 0938-7994 .- 1432-1084. ; 24:9, s. 2279-2291
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE:To measure the test-retest reproducibility of an automated system for quantifying whole body and compartmental muscle volumes using wide bore 3 T MRI.MATERIALS AND METHODS:Thirty volunteers stratified by body mass index underwent whole body 3 T MRI, two-point Dixon sequences, on two separate occasions. Water-fat separation was performed, with automated segmentation of whole body, torso, upper and lower leg volumes, and manually segmented lower leg muscle volumes.RESULTS:Mean automated total body muscle volume was 19·32 L (SD9·1) and 19·28 L (SD9·12) for first and second acquisitions (Intraclass correlation coefficient (ICC) = 1·0, 95 % level of agreement -0·32-0·2 L). ICC for all automated test-retest muscle volumes were almost perfect (0·99-1·0) with 95 % levels of agreement 1.8-6.6 % of mean volume. Automated muscle volume measurements correlate closely with manual quantification (right lower leg: manual 1·68 L (2SD0·6) compared to automated 1·64 L (2SD 0·6), left lower leg: manual 1·69 L (2SD 0·64) compared to automated 1·63 L (SD0·61), correlation coefficients for automated and manual segmentation were 0·94-0·96).CONCLUSION:Fully automated whole body and compartmental muscle volume quantification can be achieved rapidly on a 3 T wide bore system with very low margins of error, excellent test-retest reliability and excellent correlation to manual segmentation in the lower leg.KEY POINTS:• Sarcopaenia is an important reversible complication of a number of diseases. • Manual quantification of muscle volume is time-consuming and expensive. • Muscles can be imaged using in and out of phase MRI. • Automated atlas-based segmentation can identify muscle groups. • Automated muscle volume segmentation is reproducible and can replace manual measurements.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Segmentation
MRI
Muscle
Volume
Reliablity

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy