Comparing 2D and 3D Magnetic Resonance Elastography Techniques in a Clinical Setting: Initial Experiences

Forsgren, Mikael (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Radiofysikavdelningen US
Norén, Bengt (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Röntgenkliniken i Linköping
Kihlberg, Johan, 1970- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Röntgenkliniken i Linköping
visa fler...
Dahlqvist Leinhard, Olof (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Radiofysikavdelningen US
Kechagias, Stergios (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för kardiovaskulär medicin,Hälsouniversitetet,Magtarmmedicinska kliniken
Lundberg, Peter (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Radiofysikavdelningen US,Röntgenkliniken i Linköping
visa färre...
 (creator_code:org_t)
2014
2014
Engelska.
  • Konferensbidrag (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • Purpose: It has been shown that liver fibrosis, and even cirrhosis, may be reversible in humans. For this reason there is a great need for the imminent introduction of non-invasive and clinically useful methods in order to monitor fibrosis in patients [1, 2]. A body of evidence points to the fact that MRE is a highly useful candidate towards this end [3]. However, before using such techniques more widely, it is important to verify that comparable physical measures are provided by alternative and clinically relevant MRE approaches. The aim of this pilot study was to compare 2D and 3D MRE, also known as MR Rheology, using a commercially available 2D system, with an acoustic transducer, and 3D MRE research system, with an electromagnetic transducer, with respect to liver stiffness and elasticity in patients with diffuse or suspected diffuse liver disease. Materials and Methods: Seven patients, referred to our hospital for evaluation of elevated serum alanine aminotransferase (ALT) and/or alkaline phosphatase (ALP) levels but without signs of cirrhosis on physical examination, were recruited from a previous study [4], and examined in the course of one day. Fibrosis staging from prior biopsy were gained from [4], see Table 1. The 3D MRE method included an active electromagnetic transducer generating waves at 56 Hz, and a 1.5 T Philips Achieva MR-scanner, with a phased array body coil (Sense TorsoXL, all 16 coil elements), GRE sequence parameters include; FOV = 320x256 mm2, matrix = 80x38, slice thickness = 4 mm, # slices = 9, FA = 15°, TR = 112 ms, TE = 9.21 ms, SENSE = 2. The 2D MRE method included a passive acoustic transducer generating waves at 60 Hz, and a 1.5 T GE 450W MR-scanner, with a phased array body coil (HD8 Torso, all 8 coil elements), GRE sequence parameters include; FOV = 440x440 mm2, matrix = 256x64, slice thickness = 10 mm, # slices = 4, FA = 30°, TR = 50 ms, TE = 21.7 ms, ASSET = 2. The transducers were on both systems placed on the anterior chest wall to the right of xiphoid process (patient in a supine position), the time between each MRE acquisition was dependent on how long it took to transfer the patient between the two MR systems in the hospital (<10 min) A region of interest (ROI) was placed in an appropriate single 10 mm slice acquired using the GE MR-scanner. A corresponding ROI for the Philips system, covering the same anatomical region, was placed over three slices (4 mm thickness each). This yielded a total cranio-caudal coverage of the ROIs equal to 10 mm (on the GE data) and 12 mm (on the Philips data). The mean and standard deviations of the stiffness (GE), elasticity (Philips) and Gabs,Elastic (Philips) were calculated. Gabs,Elastic is the absolute value of the shear modulus, which in principle is equivalent to the viscoelastic property, shear stiffness. In the 3D method the shear waves were obtained by applying the curl operator and using the Voigt rheological model to obtain shear elasticity maps [5, 6]. In the 2D method the GE system provided the stiffness maps. Statistics was performed using Mathematica 9. ROI drawing and quantification of the data from the GE system was performed using Sectra PACS IDS7, and ROI drawing and quantification of the data from the Philips system was performed using a custom software package implemented in ROOT, generously provided by R. Sinkus (Kings College, London, UK). Results: The measured values are presented in Table 1. Both elasticity and Gabs,Elastic correlates well with the stiffness measurement carried out in the GE system (Fig. 1), as was shown by the elasticity and stiffness correlation R2 = 0.96 (P < 0.001) slope = 1.08 (P < 0.001), intercept = 0.61 kPa (P = 0.08), Gabs,Elastic and stiffness correlation R2 = 0.96 (P < 0.001), slope = 0.95 (P< 0.001) intercept = 0.28 kPa (P = 0.43)

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

vet (ämneskategori)
kon (ämneskategori)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy