The shape of the proximal isovelocity surface area varies with regurgitant orifice size and distance from orifice : computer simulation and model experiments with color M-mode technique.

Barclay, Susan A (författare)
Eidenvall, Lars (författare)
Karlsson, Matts (författare)
visa fler...
Andersson, Gunnar (författare)
Xiong, Changsheng (författare)
Ask, Per (författare)
Linköpings universitet,Fysiologisk mätteknik,Tekniska högskolan
Loyd, Dan (författare)
Linköpings universitet,Mekanisk värmeteori och strömningslära,Tekniska högskolan
Wranne, Bengt (författare)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken US
visa färre...
 (creator_code:org_t)
1993
1993
Engelska.
Ingår i: Journal of the American Society of Echocardiography. - 0894-7317 .- 1097-6795. ; 6:4, s. 433-445
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The hemispheric proximal isovelocity surface area method for quantification of mitral regurgitant flow (i.e., Qc = 2 pi r2v), where 2 pi r2 is the surface area and v is the velocity at radius r, was investigated as distance from the orifice was increased. Computer simulations and steady flow model experiments were performed for orifices of 4, 6, and 8 mm. Flow rates derived from the centerline velocity and hemispheric assumption were compared with true flow rates. Proximal isovelocity surface area shape varied as distance from each orifice was increased and could only be approximated from the hemispheric equation when a certain distance was exceeded: > 7, > 10, and > 12 mm for the 4, 6, and 8 mm orifices, respectively. Prediction of relative error showed that the best radial zone at which to make measurements was 5 to 9, 6 to 14 and 7 to 17 mm for the 4, 6, and 8 mm orifices, respectively. Although effects of a nonhemispheric shape could be compensated for by use of a correction factor, a radius of 8 to 9 mm can be recommended without the use of a correction factor over all orifices studied if a deviation in calculated as compared with true flow of 15% is considered acceptable. These measurements therefore have implications for the technique in clinical practice.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy