A generic TG-186 shielded applicator for commissioning model-based dose calculation algorithms for high-dose-rate Ir-192 brachytherapy

Ma, Yunzhi (författare)
CHU Quebec, Canada; University of Laval, Canada; University of Laval, Canada
Vijande, Javier (författare)
University of Valencia, Spain; IFIC, Spain
Ballester, Facundo (författare)
University of Valencia, Spain
visa fler...
Carlsson Tedgren, Åsa (författare)
Karolinska Institutet,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US,Karolinska University Hospital, Sweden
Granero, Domingo (författare)
Hospital Gen University, Spain
Haworth, Annette (författare)
University of Sydney, Australia
Mourtada, Firas (författare)
Christiana Care Health Syst, DE 19713 USA; Christiana Care Health Syst, DE 19713 USA
Paiva Fonseca, Gabriel (författare)
Maastricht University, Netherlands
Zourari, Kyveli (författare)
University of Athens, Greece
Papagiannis, Panagiotis (författare)
University of Athens, Greece,Carleton University, Canada
Rivard, Mark J. (författare)
Tufts University, MA 02111 USA
Siebert, Frank-Andre (författare)
University Hospital Schleswig Holstein, Germany
Sloboda, Ron S. (författare)
Cross Cancer Institute, Canada; University of Alberta, Canada
Smith, Ryan (författare)
Alfred Hospital, Australia
Chamberland, Marc J. P. (författare)
Carleton University, Canada
Thomson, Rowan M. (författare)
Carleton University, Canada
Verhaegen, Frank (författare)
Maastricht University, Netherlands
Beaulieu, Luc (författare)
CHU Quebec, Canada; University of Laval, Canada; University of Laval, Canada
visa färre...
 (creator_code:org_t)
2017-10-19
2017
Engelska.
Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405 .- 2473-4209. ; 44:11, s. 5961-5976
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PurposeA joint working group was created by the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the Australasian Brachytherapy Group (ABG) with the charge, among others, to develop a set of well-defined test case plans and perform calculations and comparisons with model-based dose calculation algorithms (MBDCAs). Its main goal is to facilitate a smooth transition from the AAPM Task Group No. 43 (TG-43) dose calculation formalism, widely being used in clinical practice for brachytherapy, to the one proposed by Task Group No. 186 (TG-186) for MBDCAs. To do so, in this work a hypothetical, generic high-dose rate (HDR) Ir-192 shielded applicator has been designed and benchmarked. MethodsA generic HDR Ir-192 shielded applicator was designed based on three commercially available gynecological applicators as well as a virtual cubic water phantom that can be imported into any DICOM-RT compatible treatment planning system (TPS). The absorbed dose distribution around the applicator with the TG-186 Ir-192 source located at one dwell position at its center was computed using two commercial TPSs incorporating MBDCAs (Oncentra((R)) Brachy with Advanced Collapsed-cone Engine, ACE, and BrachyVision ACUROS) and state-of-the-art Monte Carlo (MC) codes, including ALGEBRA, BrachyDose, egs_brachy, Geant4, MCNP6, and Penelope2008. TPS-based volumetric dose distributions for the previously reported source centered in water and source displaced test cases, and the new source centered in applicator test case, were analyzed here using the MCNP6 dose distribution as a reference. Volumetric dose comparisons of TPS results against results for the other MC codes were also performed. Distributions of local and global dose difference ratios are reported. ResultsThe local dose differences among MC codes are comparable to the statistical uncertainties of the reference datasets for the source centered in water and source displaced test cases and for the clinically relevant part of the unshielded volume in the source centered in applicator case. Larger local differences appear in the shielded volume or at large distances. Considering clinically relevant regions, global dose differences are smaller than the local ones. The most disadvantageous case for the MBDCAs is the one including the shielded applicator. In this case, ACUROS agrees with MC within [-4.2%, +4.2%] for the majority of voxels (95%) while presenting dose differences within [-0.12%, +0.12%] of the dose at a clinically relevant reference point. For ACE, 95% of the total volume presents differences with respect to MC in the range [-1.7%, +0.4%] of the dose at the reference point. ConclusionsThe combination of the generic source and generic shielded applicator, together with the previously developed test cases and reference datasets (available in the Brachytherapy Source Registry), lay a solid foundation in supporting uniform commissioning procedures and direct comparisons among treatment planning systems for HDR Ir-192 brachytherapy.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Ir-192; HDR brachytherapy; model based dose calculation; Monte Carlo methods; shielded applicator; TG-186

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy