Characterization of Cardiac Flow in Heart Disease Patients by CFD and 4D Flow MRI

Lantz, Jonas (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Gupta, Vikas (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Henriksson, Lilian (författare)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
visa fler...
Karlsson, Matts (författare)
Linköpings universitet,Mekanisk värmeteori och strömningslära,Tekniska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Persson, Anders (författare)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
Carlhäll, Carljohan (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Fysiologiska kliniken US
Ebbers, Tino (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Fysiologiska kliniken US
visa färre...
 (creator_code:org_t)
American Physical Society, 2017
2017
Engelska.
Ingår i: Bulletin of the Amerian Physcial Society. - : American Physical Society.
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • In this study, cardiac blood flow was simulated using Computational Fluid Dynamics and compared to in vivo flow measurements by 4D Flow MRI. In total, nine patients with various heart diseases were studied. Geometry and heart wall motion for the simulations were obtained from clinical CT measurements, with 0.3x0.3x0.3 mm spatial resolution and 20 time frames covering one heartbeat. The CFD simulations included pulmonary veins, left atrium and ventricle, mitral and aortic valve, and ascending aorta. Mesh sizes were on the order of 6-16 million cells, depending on the size of the heart, in order to resolve both papillary muscles and trabeculae. The computed flow field agreed visually very well with 4D Flow MRI, with characteristic vortices and flow structures seen in both techniques. Regression analysis showed that peak flow rate as well as stroke volume had an excellent agreement for the two techniques. We demonstrated the feasibility, and more importantly, fidelity of cardiac flow simulations by comparing CFD results to in vivo measurements. Both qualitative and quantitative results agreed well with the 4D Flow MRI measurements. Also, the developed simulation methodology enables “what if” scenarios, such as optimization of valve replacement and other surgical procedures.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
kon (ämneskategori)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy