SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Ballester Facundo)
 

Sökning: WFRF:(Ballester Facundo) > GEC-ESTRO ACROP rec...

GEC-ESTRO ACROP recommendations on calibration and traceability of LE-LDR photon-emitting brachytherapy sources at the hospital level

Perez-Calatayud, Jose (författare)
Univ and Polytech La Fe Hosp, Spain; IIS La Fe UV, Spain
Ballester, Facundo (författare)
IIS La Fe UV, Spain; Univ Valencia, Spain
Carlsson Tedgren, Åsa (författare)
Karolinska Institutet,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik,Karolinska Univ Hosp, Sweden; Karolinska Inst, Sweden
visa fler...
Rijnders, Alex (författare)
Europe Hosp, Belgium
Rivard, Mark J. (författare)
Brown Univ, RI 02912 USA
Andrassy, Michael (författare)
RandD Brachytherapy Eckert and Ziegler BEBIG, Germany
Niatsetski, Yury (författare)
RandD Elekta Brachytherapy, Netherlands
Schneider, Thorsten (författare)
PTB, Germany
Siebert, Frank-Andre (författare)
UK S H, Germany
visa färre...
 (creator_code:org_t)
ELSEVIER IRELAND LTD, 2019
2019
Engelska.
Ingår i: Radiotherapy and Oncology. - : ELSEVIER IRELAND LTD. - 0167-8140 .- 1879-0887. ; 135, s. 120-129
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Prostate brachytherapy treatment using permanent implantation of low-energy (LE) low-dose rate (LDR) sources is successfully and widely applied in Europe. In addition, seeds are used in other tumour sites, such as ophthalmic tumours, implanted temporarily. The calibration issues for LE-LDR photon emitting sources are specific and different from other sources used in brachytherapy. In this report, the BRAPHYQS (BRAchytherapy PHYsics Quality assurance System) working group of GEC-ESTRO, has developed the present recommendations to assure harmonized and high-quality seed calibration in European clinics. There are practical aspects for which a clarification/procedure is needed, including aspects not specifically accounted for in currently existing AAPM and ESTRO societal recommendations. The aim of this report has been to provide a European wide standard in LE-LDR source calibration at end-user level, in order to keep brachytherapy treatments with high safety and quality levels. The recommendations herein reflect the guidance to the ESTRO brachytherapy users and describe the procedures in a clinic or hospital to ensure the correct calibration of LE-LDR seeds. (C) 2019 The Authors. Published by Elsevier B.V.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Brachytherapy; LE-LDR; Seeds; Calibration

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy