Cutaneous bioengineering instrumentation standardization : the Tissue Viability Imager

Nilsson, Gert (author)
Linköpings universitet,Biomedicinsk instrumentteknik,Tekniska högskolan
Zhai, Hongbo (author)
P Chan, Heidi (author)
University of California
show more...
Farahmand, Sara (author)
University of California
Maibach, Howard I (author)
University of California
show less...
 (creator_code:org_t)
Wiley, 2009
2009
English.
In: Skin research and technology. - : Wiley. - 0909-752X .- 1600-0846. ; 15:1, s. 6-13
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Tissue Viability Imaging (TiVi) is a new bioengineering technology intended for remote two-dimensional mapping of skin red blood cell concentration (RBCconc). Before use in the laboratory, work-site and dermatology clinic, critical performance parameters of this emerging technology require careful evaluation. To assess short- and long-term stability, image uniformity, distance and image size dependence, ambient light and curvature influence in a production batch of Tissue Viability Imagers. Four Tissue Viability Imagers from the same production batch were evaluated at two laboratories (one industrial and one dermatological) with respect to critical parameter performance. The average systematic drift in sensitivity over time was 0.27% and < 1.02% for all four units tested. Difference in sensitivity between units was limited to 4.1% and was due to offset rather than gain deviation. Spatial variation in image uniformity was below 3.08% and 1.93% in the corners and centre of an individual image, respectively. This spatial variation could be further reduced to 0.25% and 0.13%, respectively by image normalization. Ambient light from a 40 W bulb or a 11 W fluorescent light source at a distance of 50-60 cm above the object, reduced the recorded values by about 10%, while the camera to object distance and image size had no detectable influence on sensitivity. Curved objects, such as human forearm, demonstrated an edge effect limited to below 10%. The critical TiVi performance parameters evaluated proved stable in relation to expected variations in skin RBCconc over time. Calibration by way of a two-point method may reduce differences in sensitivity between instruments to further facilitate inter-laboratory comparison of results.

Keyword

Tissue Viability Imaging
skin red blood cell concentration
spectroscopy
instrument performance
skin bioengineering
MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view