Comparison of Venous and Capillary Sampling in Oral Glucose Testing for the Diagnosis of Gestational Diabetes Mellitus: A Diagnostic Accuracy Cross-Sectional Study Using Accu-Chek Inform II

Nevander, Sofia (author)
Linköpings universitet,Avdelningen för barns och kvinnors hälsa,Medicinska fakulteten,Region Östergötland, Kvinnokliniken ViN
Landberg, Eva (author)
Linköpings universitet,Avdelningen för cellbiologi,Medicinska fakulteten,Region Östergötland, Klinisk kemi
Blomberg, Marie (author)
Linköpings universitet,Avdelningen för barns och kvinnors hälsa,Medicinska fakulteten,Region Östergötland, Kvinnokliniken US
show more...
Ekman, Bertil, 1958- (author)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Region Östergötland, Endokrinmedicinska kliniken
Lilliecreutz, Caroline (author)
Linköpings universitet,Avdelningen för barns och kvinnors hälsa,Medicinska fakulteten,Region Östergötland, Kvinnokliniken US
show less...
 (creator_code:org_t)
2020-11-26
2020
English.
In: Diagnostics (Basel). - : MDPI. - 2075-4418. ; 10:12
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Gestational diabetes mellitus (GDM) is a common complication with negative impacts on mother and child. The primary aim of this study was to examine whether plasma glucose cutoffs for GDM diagnosis based on venous sampling can be replaced by cutoffs based on capillary sampling. A prospective cross-sectional study was performed at an antenatal care clinic including 175 pregnant women undergoing an oral glucose tolerance test (OGTT). Duplicate samples were collected by capillary and venous puncture while fasting and 1 h and 2 h after an OGTT. Both samples were analyzed on Accu-Chek Inform II. The cutoffs for a GDM diagnosis using capillary samples were corrected from 5.1 to 5.3 mmol/L for the fasting sample, from 10.0 to 11.1 mmol/L for the 1 h sample, and from 8.5 to 9.4 mmol/L for the 2-h sample using half of the dataset. Applying these cutoffs to the remaining dataset resulted in a sensitivity, specificity, and accuracy of 85.0%, 95.0%, and 90.3%, respectively, with a positive predictive value (PPV) of 83%, an negative predictive value (NPV) of 96%, and a positive negative likelihood ratio (LHR) of 16.4 using capillary sampling for the GDM diagnosis at fasting and 2-h after. Corrected cutoffs and capillary samples can be used for the diagnosis of GDM with maintained diagnostic accuracy using Accu-Chek Inform II.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

antenatal care; gestational diabetes mellitus; pregnancy; OGTT; fasting glucose; capillary sampling

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view