Optimal Selection of Base Materials for Accurate Dual-Energy Computed Tomography : Comparison Between the Alvarez–Macovski Method and DIRA

Magnusson, Maria, 1961- (författare)
Linköpings universitet,Datorseende,Tekniska fakulteten,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Alm Carlsson, Gudrun, 1939- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Sandborg, Michael, 1961- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik
visa fler...
Carlsson Tedgren, Åsa, 1968- (författare)
Karolinska Institutet,Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik,Department of Medical Radiation Physics and Nuclear Medicine; Karolinska University Hospital, Stockholm, Sweden
Malusek, Alexandr, 1968- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
visa färre...
 (creator_code:org_t)
2021-07-08
2021
Engelska.
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press. - 0144-8420 .- 1742-3406. ; 195:3-4, s. 218-224
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The choice of the material base to which the material decomposition is performed in dual-energy computed tomography may affect the quality of reconstructed images. The aim of this work is to investigate how the commonly used bases (water, bone), (water, iodine) and (photoelectric effect, Compton scattering) affect the reconstructed linear attenuation coefficient in the case of the Alvarez–Macovski method. The performance of this method is also compared with the performance of the Dual-energy Iterative Reconstruction Algorithm (DIRA). In both cases, the study is performed using simulations. The results show that the Alvarez–Macovski method produced artefacts when iodine was present in the phantom together with human tissues since this method can only work with one doublet. It was shown that these artefacts could be avoided with DIRA using the (water, bone) doublet for tissues and the (water, iodine) doublet for the iodine solution.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Public Health
Environmental and Occupational Health
Radiology Nuclear Medicine and imaging
General Medicine
Radiation
Radiological and Ultrasound Technology

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy