Doppler flow measurement using surface integration of velocity vectors (Sivv) : in vitro validation

Chew, Michelle (författare)
Departments of Anaesthesia and Intensive Care, Aarhus, Denmark
Brandberg, Joakim, 1967- (författare)
Linköpings universitet,Institutionen för medicinsk teknik,Tekniska högskolan
Canard, Peter (författare)
Institute of Experimental Clinical Research, Skejby Sygehus, Aarhus University Hospital, Aarhus, Denmark
visa fler...
Sloth, Erik (författare)
Departments of Anaesthesia and Intensive Care, Aarhus, Denmark
Ask, Per, 1950- (författare)
Linköpings universitet,Institutionen för medicinsk teknik,Tekniska högskolan,Fysiologisk mätteknik
Hasenkam, Michael (författare)
Department of Cardiothoracic and Vascular Surgery, Aarhus, Denmark
visa färre...
 (creator_code:org_t)
2000
2000
Engelska.
Ingår i: Ultrasound in Medicine and Biology. - 0301-5629 .- 1879-291X. ; 26:2, s. 255-262
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Blood flow measurement using an improved surface integration of velocity vectors (SIVV) technique was tested in in vitro phantoms. SIVV was compared with true flow (12–116 mL/s) in a steady-state model using two angles of insonation (45° and 60°) and two vessel sizes (internal diameter = 11 and 19 mm). Repeatability of the method was tested at various flow rates for each angle of insonation and vessel. In a univentricular pulsatile model, SIVV flow measured at the mitral inlet was compared to true flow (29–61 mL/s). Correlation was excellent for the 19-mm vessel (r2= 0.99). There was a systematic bias but close limits of agreement (mean ± 2 SD = −24.1% ± 7.6% at 45 °; +16.4% ± 11.0% at 60 °). Using the 11-mm vessel, a quadratic relationship was demonstrated between between SIVV and true flow (r2 = 0.98–0.99), regardless of the angle of insonation. In the pulsatile system, good agreement and correlation were shown (r2 = 0.94, mean ± 2 SD = −4.7 ± 10.1%). The coefficients of variation for repeated SIVV measurements ranged from 0.9% to 10.3%. This method demonstrates precision and repeatability, and is potentially useful for clinical measurements.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy