  • The aim of this study was to develop a finite element model for simulation of the thermal characteristics of brain electrodes and to compare its performances with an in vitro experimental albumin model. Ten lesions were created in albumin using a monopolar electrode connected to a Leksell Neuro Generator and a computer-assisted video system was used to determine the size of the generated lesions. A finite element model was set up of the in vitro experiments using the same thermal properties. With a very simple heat source applied to the finite element model in the proximity of the upper part of the tip, a good agreement (no deviations in width and distance from tip but a deviation in length of −1.6 mm) with the in vitro experiments (width 4.6±0.1 mm and length 7.4±0.1 mm) was achieved when comparing the outline of the lesion. In addition, a gelatinous albumin-model was set up and compared to computer simulations resulting in deviations in width of −0.4 mm, length of −2.2 mm and distance from the tip of −0.1 mm. Hence, the utilisation of finite element model simulations may be a useful complement to in-vitro experiments.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy