Evaluation of Enhanced High-Resolution Laser Doppler Imaging in an in Vitro Tube Model with the Aim of Assessing Blood Flow in Separate Microvessels

Lindén, Maria, 1965- (författare)
Linköpings universitet,Institutionen för medicinsk teknik,Hälsouniversitetet
Golster, Helena (författare)
Linköpings universitet,Hand och plastikkirurgi,Anestesiologi med intensivvård,Hälsouniversitetet
Bertuglia, Silvia (författare)
CNR Institute of Clinical Physiology, Via Trieste, Pisa, Italy
visa fler...
Colantuoni, Antonio (författare)
CNR Institute of Clinical Physiology, Via Trieste, Pisa, Italy
Sjöberg, Folke, 1956- (författare)
Linköpings universitet,Hand och plastikkirurgi,Anestesiologi med intensivvård,Hälsouniversitetet
Nilsson, Gert, 1947- (författare)
Linköpings universitet,Institutionen för medicinsk teknik,Hälsouniversitetet
visa färre...
 (creator_code:org_t)
Elsevier BV, 1998
1998
Engelska.
Ingår i: Microvascular Research. - : Elsevier BV. - 0026-2862 .- 1095-9319. ; 56:3, s. 261-270
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • An enhanced high-resolution laser Doppler imaging (EHR-LDI) technique intended for visualization of separate microvessels was evaluated by use ofin vitroflow models. In EHR-LDI, a laser beam focused to a half-power diameter less than 40 μm successively scans the tissue under study in steps of 25 μm. Spatial blood flow variations within microvascular structures of 1.5 × 1.5 mm are rendered by 64 × 64 measurement sites. Individual microvessel diameters could be estimated and an average difference of 11 μm compared to microscopic measurements was obtained. For the flow algorithm used, the LDI output signal was found to scale linearly with average velocity (0–3.5 mm/s) when a plastic tube of inner diameter 175 μm was perfused with human blood (correlation coefficient 0.99). The LDI output signal was further found insensitive to hematocrit variations in the range 16–44%. Due to the limited laser light penetration in blood, a reduction in the LDI output signal was observed as the inner tube diameters were successively changed from 280 to 1400 μm.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy