SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Ritter A.)
 

Sökning: WFRF:(Ritter A.) > (2005-2009) > Guidelines for samp...

Guidelines for sampling, measuring and reporting ionized magnesium in undiluted serum, plasma or blood : International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)

Ben, Rayana M.C. (författare)
Ben Rayana, M.C., Laboratory of Clinical Chemistry, National Institute of Nutrition, Bab Saadoun, Tunis, Tunisia
Burnett, R.W. (författare)
Department of Pathology, Hartford Hospital, Hartford, CT, United States
Covington, A.K. (författare)
Department of Chemistry, University of Newcastle upon Tyne, Newcastle upon Tyne, United Kingdom
visa fler...
D'Orazio, P. (författare)
Instrumentation Laboratory, Lexington, MA, United States
Fogh-Andersen, N. (författare)
Laboratory of Clinical Chemistry, Herlev Hospital, Herlev, Denmark
Jacobs, E. (författare)
Clinical Laboratory Program, Wadsworth Center, NY State Deparmtent of Health, Albany, NY, United States
Kulpmann, W.R. (författare)
Külpmann, W.R., Klinische Chemie, Medizinische Hochschule, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
Kuwa, K. (författare)
Institute of Clinical Medicine, University of Tsukuba, Tsukuba, Japan
Larsson, Lasse (författare)
Linköpings universitet,Hälsouniversitetet,Klinisk kemi
Lewenstam, A. (författare)
Center ProSens, Åbo Akademi University, Åbo-Turku, Finland
Maas, A.H.J. (författare)
Eurotrol bv, Wageningen, Netherlands
Mager, G. (författare)
Fresenius Medical Care Deutschland GmbH, Bad Homburg, Germany
Naskalski, J.H.J. (författare)
Department of Clinical Biochemistry, Colleguim Medicum, Jagiellonian University, Krakow, Poland
Okorodudu, A.O. (författare)
Department of Pathology, John Sealy Hospital, Galveston, TX, United States
Ritter, C. (författare)
Roche Diagnostics GmbH, Graz, Austria
St, John A. (författare)
St John, A., ARC Consulting, MT Lawley, WA, Australia
visa färre...
Ben Rayana, MC., Laboratory of Clinical Chemistry, National Institute of Nutrition, Bab Saadoun, Tunis, Tunisia Department of Pathology, Hartford Hospital, Hartford, CT, United States (creator_code:org_t)
2005
2005
Engelska.
Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 43:5, s. 564-569
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • All analyzers with ion-selective electrodes for ionized magnesium (iMg) should yield comparable and unbiased results. The prerequisite to achieve this goal is to reach consensus on sampling, measurement and reporting. The recommended guidelines for sampling, measurement and reporting iMg in plasma ("plasma" refers to circulating plasma and the forms in which it is sampled: the plasma phase of anticoagulated whole blood, plasma separated from blood cells, or serum) or blood, referring to the substance concentration of iMg in the calibrants, will provide results for iMg that are approximately 3% greater than its true concentration, and 4% less than its true molality. Binding of magnesium to proteins and ligands in plasma and blood is pH-dependent. Therefore, pH should be simultaneously measured to allow adjustment of iMg concentration to pH 7.4. The substance concentration of iMg may be physiologically and consequently clinically more relevant than the substance concentration of total magnesium. © 2005 by Walter de Gruyter.

Nyckelord

Erythrocyte
Heparin
Influence factors
Interference
Ion-selective electrode
Ionized calcium
Ionized magnesium
Lipophilic anions
Liquid junction potential
pH
Protein
Sampling
Silicone
Sodium
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy