IFCC guideline for sampling, measuring and reporting ionized magnesium in plasma

Ben Rayana, Mohammed C. (författare)
Natl Inst Nutr, Clin Chem Lab, Tunis, Tunisia
Burnett, Robert W. (författare)
Hartford Hosp, Dept Pathol, Hartford, CT 06115 USA
Covington, Arthur K. (författare)
Univ Newcastle Upon Tyne, Dept Chem, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
visa fler...
D'Orazio, Paul (författare)
Instrumentat Lab, Lexington, MA USA
Fogh-Andersen, Niels (författare)
Herlev Hosp, Dept Clin Biochem, Clin Chem Lab, DK-2730 Herlev, Denmark
Jacobs, Ellis (författare)
NY State Dept Hlth, Wadsworth Ctr, Clin Lab Program, Albany, NY USA
Kulpmann, Wolf R. (författare)
Med Hochschule, Klin Chem, Hannover, Germany
Kuwa, Katsuhiko (författare)
Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 305, Japan
Larsson, Lasse (författare)
Linköpings universitet,Hälsouniversitetet,Klinisk kemi
Lewenstam, Andrzej (författare)
Abo Akad Univ, Ctr ProSens, Turku, Finland
Maas, Anton H. J. (författare)
Eurotrol Bv, Ede, Netherlands
Mager, Gerhard (författare)
Fresenius Med Care, Bad Homburg, Germany
Naskalski, Jerzy W. (författare)
Jagiellonian Univ, Coll Med, Clin Biochem, Krakow, Poland
visa färre...
 (creator_code:org_t)
2008
2008
Engelska.
Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 46:1, s. 21-26
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Analyzers with ion-selective electrodes (ISEs) for ionized magnesium (iMg) should yield comparable and unbiased results for iMg. This IFCC guideline on sampling, measuring and reporting iMg in plasma provides a prerequisite to achieve this goal [in this document, "plasma" refers to circulating plasma and the forms in which it is sampled, namely the plasma phase of anticoagulated whole blood (or "blood"), plasma separated from blood cells, or serum]. The guideline recommends measuring and reporting ionized magnesium as a substance concentration relative to the substance concentration of magnesium in primary aqueous calibrants with magnesium, sodium, and calcium chloride of physiological ionic strength. The recommended name is "the concentration of ionized magnesium in plasma". Based on this guideline, results will be approximately 3% higher than the true substance concentration and 4% lower than the true molality in plasma. Calcium ions interfere with all current magnesium ion-selective electrodes (Mg-ISEs), and thus it is necessary to determine both ions simultaneously in each sample and correct the result for Ca2+ interference. Binding of Mg in plasma is pH-dependent. Therefore, pH should be measured simultaneously with iMg to allow adjustment of the result to pH 7.4. The concentration of iMg in plasma may be physiologically and clinically more relevant than the concentration of total magnesium. Furthermore, blood-gas analyzers or instruments for point-of-care testing are able to measure plasma iMg using whole blood (with intact blood cells) as the sample, minimizing turnaround time compared to serum and plasma, which require removal of blood cells.

Nyckelord

erythrocyte effect
heparin
influence factors
interference
ionized calcium
ionized magnesium
ion-selective electrode
lipophilic anions
liquid junction potential
pH
protein
sampling
silicone
sodium
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy