A technique based on laser Doppler flowmetry and photoplethysmography for simultaneously monitoring blood flow at different tissue depths

Hagblad, Jimmie (författare)
Mälardalens högskola,Akademin för innovation, design och teknik,Mälardalen University
Lindberg, Lars-Göran (författare)
Mälardalens högskola,Linköpings universitet,Fysiologisk mätteknik,Tekniska högskolan,Akademin för innovation, design och teknik
Kaisdotter Andersson (Jonsson), Annika (författare)
Mälardalens högskola,Akademin för innovation, design och teknik,Mälardalen University
visa fler...
Bergstrand, Sara (författare)
Mälardalens högskola,Linköpings universitet,Omvårdnad,Hälsouniversitetet,Akademin för innovation, design och teknik
Lindgren, Margareta (författare)
Mälardalens högskola,Linköpings universitet,Omvårdnad,Hälsouniversitetet,Akademin för innovation, design och teknik,Mälardalen University
Ek, Anna-Christina (författare)
Mälardalens högskola,Östergötlands Läns Landsting,Linköpings universitet,Omvårdnad,Hälsouniversitetet,Akutkliniken,Akademin för innovation, design och teknik
Folke, Mia (författare)
Mälardalens högskola,Akademin för innovation, design och teknik,Mälardalen University
Lindén, Maria (författare)
Mälardalens högskola,Akademin för innovation, design och teknik
visa färre...
 (creator_code:org_t)
2010-01-28
2010
Engelska.
Ingår i: Medical and Biological Engineering and Computing. - : Springer Science Business Media. - 0140-0118 .- 1741-0444. ; 48:5, s. 415-422
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of this study was to validate a non-invasive optical probe for simultaneous blood flow measurement at different vascular depths combining three photoplethysmography (PPG) channels and laser Doppler flowmeter (LDF). Wavelengths of the PPG were near-infrared 810 nm with source-to-detector separation of 10 and 25 mm, and green 560 nm with source-to-detector separation of 4 mm. The probe is intended for clinical studies of pressure ulcer aetiology. The probe was placed over the trapezius muscle, and depths from the skin to the trapezius muscle were measured using ultrasound and varied between 3.8 and 23 mm in the 11 subjects included. A provocation procedure inducing a local enhancement of blood flow in the trapezius muscle was used. Blood flows at rest and post-exercise were compared. It can be concluded that this probe is useful as a tool for discriminating between blood flows at different vascular tissue depths. The vascular depths reached for the different channels in this study were at least 23 mm for the near-infrared PPG channel (source-to-detector separation 25 mm), 10-15 mm for the near-infrared PPG channel (separation 10 mm), and shallower than 4 mm for both the green PPG channel (separation 4 mm) and LDF.

Nyckelord

PPG
LDF
Perfusion
Vascular depth
Pressure ulcer
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy