Distributions of scatter-to-primary ratios in chest PA radiography using grid or air gap for scatter-rejection
-
- Ullman, Gustav (author)
- Linköpings universitet,Radiofysik,Hälsouniversitetet
-
- Sandborg, Michael (author)
- Östergötlands Läns Landsting,Linköpings universitet,Radiofysik,Hälsouniversitetet,Radiofysikavdelningen
-
- Dance, David R (author)
- Department of Physics, The Royal Marsden NHS Trust, UK
-
show more...
-
- Hunt, Roger (author)
- Department of Physics, The Royal Marsden NHS Trust, UK
-
- Alm Carlsson, Gudrun (author)
- Östergötlands Läns Landsting,Linköpings universitet,Radiofysik,Hälsouniversitetet,Radiofysikavdelningen
-
show less...
-
(creator_code:org_t)
- Linköping : Linköping University Electronic Press, 2004
- English 39 s.
-
Series: Report / Institutionen för radiofysik, Universitetet i Linköping, 1998- ; 100
- Related links:
-
https://liu.diva-por... (primary) (Raw object)
-
show more...
-
https://urn.kb.se/re...
-
show less...
Abstract
Subject headings
Close
- The aim for this work was to calculate distributions of scatter-to-primary ratios (εs/εp) and signal-to-noise ratios per pixel (SNRp) for a large set of imaging systems with either grid or air gap for scatter rejection. Grids with ratio 8-16 and air gap length 20 and 40 cm were used. The tube voltage was varied between 90-150 kV and three patient thickness between 20-28 cm were tested in order to compare scatter-rejections techniques for different conditions. Distributions of this sort may provide useful information on how physical image quality (contrast, SNR) is distributed over the chest PA image. A Monte Carlo computer program was used for the calculations, including a model of both the patient (voxel phantom) and the imaging system. The mean value of the εs/εp is 0.39 in the hilar region and 1.72 in the lower mediastinal region. For a 28 cm patient, the corresponding values are 0.42 in the hilar region and 2.58 in the lower mediastinal region. The grid with ratio 16 is the most efficient scatter rejection technique in all regions except the hilar region. In the hilar region, the most efficient technique is the 40 cm air gap.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Keyword
- Radiological physics
- Radiofysik
Publication and Content Type
- vet (subject category)
- rap (subject category)
To the university's database