Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor study

Rolf, Marijn P (author)
Vrije University Amsterdam Medical Centre
Hofman, Mark B M (author)
Vrije University Amsterdam Medical Centre
Gatehouse, Pete rD (author)
Royal Brompton Hospital
show more...
Markenroth-Bloch, Karin (author)
Skane University Hospital
Heymans, Martijn W (author)
Vrije University Amsterdam Medical Centre
Ebbers, Tino (author)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Fysiologi,Hälsouniversitetet,Mekanisk värmeteori och strömningslära
Graves, Martin J (author)
Cambridge University Hospital
Totman, John J (author)
Kings College London
Werner, Beat (author)
University Childrens Hospital
van Rossum, Albert C (author)
Vrije University Amsterdam Medical Centre
Kilner, Philip J (author)
Royal Brompton Hospital
Heethaar, Rob M (author)
Vrije University Amsterdam Medical Centre
show less...
 (creator_code:org_t)
Taylor and Francis / BioMed Central, 2011
2011
English.
In: JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE. - : Taylor and Francis / BioMed Central. - 1097-6647. ; 13
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Purpose: Eddy current induced velocity offsets are of concern for accuracy in cardiovascular magnetic resonance (CMR) volume flow quantification. However, currently known theoretical aspects of eddy current behavior have not led to effective guidelines for the optimization of flow quantification sequences. This study is aimed at identifying correlations between protocol parameters and the resulting velocity error in clinical CMR flow measurements in a multi-vendor study. Methods: Nine 1.5T scanners of three different types/vendors were studied. Measurements were performed on a large stationary phantom. Starting from a clinical breath-hold flow protocol, several protocol parameters were varied. Acquisitions were made in three clinically relevant orientations. Additionally, a time delay between the bipolar gradient and read-out, asymmetric versus symmetric velocity encoding, and gradient amplitude and slew rate were studied in adapted sequences as exploratory measurements beyond the protocol. Image analysis determined the worst-case offset for a typical great-vessel flow measurement. Results: The results showed a great variation in offset behavior among scanners (standard deviation among samples of 0.3, 0.4, and 0.9 cm/s for the three different scanner types), even for small changes in the protocol. Considering the absolute values, none of the tested protocol settings consistently reduced the velocity offsets below the critical level of 0.6 cm/s neither for all three orientations nor for all three scanner types. Using multilevel linear model analysis, oblique aortic and pulmonary slices showed systematic higher offsets than the transverse aortic slices (oblique aortic 0.6 cm/s, and pulmonary 1.8 cm/s higher than transverse aortic). The exploratory measurements beyond the protocol yielded some new leads for further sequence development towards reduction of velocity offsets; however those protocols were not always compatible with the time-constraints of breath-hold imaging and flow-related artefacts. Conclusions: This study showed that with current systems there was no generic protocol which resulted into acceptable flow offset values. Protocol optimization would have to be performed on a per scanner and per protocol basis. Proper optimization might make accurate (transverse) aortic flow quantification possible for most scanners. Pulmonary flow quantification would still need further (offline) correction.

Keyword

MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view