Investigation of the effect of varying scatter-to-primary ratios on nodule contrast in chest tomosynthesis

Svalkvist, Angelica (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Ullman, Gustaf (författare)
Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet,Department of Cell and Molecular Biology, Uppsala University, Uppsala, Sweden
Håkansson, Markus (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Dance, David R. (författare)
NCCPM, Royal Surrey County Hospital, Guildford , UK
Sandborg, Michael, 1961- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Centrum för medicinsk bildvetenskap och visualisering,Hälsouniversitetet,Radiofysikavdelningen US
Alm Carlsson, Gudrun (författare)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet,Radiofysikavdelningen US
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa färre...
 (creator_code:org_t)
SPIE - International Society for Optical Engineering, 2011
2011
Engelska.
Ingår i: Medical Imaging 2011. - : SPIE - International Society for Optical Engineering. - 9780819485038 ; 7961
  • Konferensbidrag (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • The primary aim of the present work was to analyze the effects of varying scatter-to-primary ratios on the appearance of simulated nodules in chest tomosynthesis section images. Monte Carlo simulations of the chest tomosynthesis system GE Definium 8000 VolumeRAD (GE Healthcare, Chalfont St. Giles, UK) were used to investigate the variation of scatter-to-primary ratios between different angular projections. The simulations were based on a voxel phantom created from CT images of an anthropomorphic chest phantom. An artificial nodule was inserted at 80 different positions in the simulated phantom images, using five different approaches for the scatter-to-primary ratios in the insertion process. One approach included individual determination of the scatter-to primary-ratio for each projection image and nodule location, while the other four approaches were using mean value, median value and zero degree projection value of the scatter-to-primary ratios at each nodule position as well as using a constant scatter-to-primary ratio of 0.5 for all nodule positions. The results indicate that the scatter-to-primary ratios vary up to a factor of 10 between the different angular tomosynthesis projections (±15°). However, the error in the resulting nodule contrast introduced by not taking all variations into account is in general smaller than 10 %.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

TSYN
XIM
Radiological physics
Radiofysik
TSYN
XIM

Publikations- och innehållstyp

vet (ämneskategori)
kon (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy