Development of a harmonised method for the profiling of amphetamines : III. Development of the gas chromatographic method

Andersson, Kjell (författare)
Swedish National Laboratory of Forensic Science
Jalava, Kaisa (författare)
National Bureau of Investigation, Finland
Lock, Eric (författare)
University of Lausanne
visa fler...
Finnon, Yvonne (författare)
University of Strathclyde, Glasgow
Huizer, Henk (författare)
Netherlands Forensic Institute
Kaa, Elisabet (författare)
University of Aarhus
Lopes, Alvaro (författare)
Policia Judiciaria, Lisbon, Portugal
Poortman-Van der Meer, Anneke (författare)
Netherlands Forensic Institute
Cole, Michael D. (författare)
University of Strathclyde, Glasgow
Dahlén, Johan (författare)
Swedish National Laboratory of Forensic Science
Sippola, Erkki (författare)
National Bureau of Investigation, Finland
visa färre...
 (creator_code:org_t)
Elsevier, 2007
2007
Engelska.
Ingår i: Forensic Science International. - : Elsevier. - 0379-0738 .- 1872-6283. ; 169:1, s. 50-63
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • This study focused on gas chromatographic analysis of target compounds found in illicit amphetamine synthesised by the Leuckart reaction, reductive amination of benzyl methyl ketone, and the nitrostyrene route. The analytical method was investigated and optimised with respect to introduction of amphetamine samples into the gas chromatograph and separation and detection of the target substances. Sample introduction using split and splitless injection was tested at different injector temperatures, and their ability to transfer the target compounds to the GC column was evaluated using cold on column injection as a reference. Taking the results from both techniques into consideration a temperature of 250 °C was considered to be the best compromise. The most efficient separation was achieved with a DB-35MS capillary column (35% diphenyl 65% dimethyl silicone; 30 m × 0.25 mm, df 0.25 μm) and an oven temperature program that started at 90 °C (1 min) and was increased by 8 °C/min to 300 °C (10 min). Reproducibility, repeatability, linearity, and limits of determination for the flame ionisation detector (FID), nitrogen phosphorous detector (NPD), and mass spectrometry (MS) in scan mode and selected ion monitoring (SIM) mode were evaluated. In addition, selectivity was studied applying FID and MS in both scan and SIM mode. It was found that reproducibility, repeatability, and limits of determination were similar for FID, NPD, and MS in scan mode. Moreover, the linearity was better when applying FID or NPD whereas the selectivity was better when utilising the MS. Finally, the introduction of target compounds to the GC column when applying injection volumes of 0.2 μl, 1 μl, 2 μl, and 4 μl with splitless injection respectively 1 μl with split injection (split ratio, 1:40) were compared. It was demonstrated that splitless injections of 1 μl, 2 μl, and 4 μl could be employed in the developed method, while split injection and splitless injections of 0.2 μl should be avoided.

Nyckelord

Amphetamine; Impurities; Profiling; Optimisation; Separation; Gas chromatography–mass spectrometry
TECHNOLOGY
TEKNIKVETENSKAP

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy