Volumetric dried blood spots for determination of phosphatidylethanol: Validation of a liquid chromatography tandem masspectrometry method and clinical application
-
Beck, Olof (författare)
-
Barroso, Miguel (författare)
-
Hermansson, Sigurd (författare)
-
visa fler...
-
Widen, Cristopher (författare)
-
Wallin, Christer (författare)
-
- Nilsson Wallmark, Cecilia, 1965 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för neurovetenskap och fysiologi, sektionen för psykiatri och neurokemi,Institute of Neuroscience and Physiology, Department of Psychiatry and Neurochemistry
-
- De Bejczy, Andrea, 1975 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för neurovetenskap och fysiologi, sektionen för psykiatri och neurokemi,Institute of Neuroscience and Physiology, Department of Psychiatry and Neurochemistry
-
visa färre...
-
(creator_code:org_t)
- 2024
- 2024
- Engelska.
-
Ingår i: DRUG TESTING AND ANALYSIS. - 1942-7603 .- 1942-7611.
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Phosphatidylethanol (PEth) measurement in whole blood samples is established as a specific alcohol biomarker with clinical and forensic applications. Establishment of dried blood spots (DBSs) as a specimen for PEth determination offers several advantages and was the focus of this work. A liquid chromatography tandem mass spectrometry method using a 96-well format for sample preparation was developed and validated. PEth was extracted from DBSs by using isopropanol containing PEth-d5 as internal standard. The blood sampling used a commercial volumetric DBS device having a phosholipase D inhibitor incorporated to stop continuous PEth formation. The method quantified PEth in the range of 0.05-10 mu mol/L, with a bias and imprecision of less than 15%. In a clinical study (n = 25) using fingerprick blood, the volumetric device offered more precise quantifications (CV 4.6%) compared with the Whatman 903 Protein Saver card device (CV 16.6%). In another clinical study (n = 48), the use of dried venous and capillary blood, and liquid venous blood was compared under real-life conditions with samples sent by postal service. The capillary and venous DBS samples gave identical results while the liquid blood gave slightly higher values. Calculation of elimination half-life (PEth 16:0/18:1) in 31 cases based on two consecutive samples with 2-9 days in between gave results (mean 6.2 days) that agree with literature but several cases with values over 10 days. In conclusion, this study demonstrates that volumetric DBS is a valid specimen for determination of PEth blood concentrations, offering several advantages. Phosphatidylethanol, a specific alcohol biomarker, measured in dried blood spots was the focus of this work. A liquid chromatography tandem mass spectrometry method was developed, validated, and applied. The blood sampling used a volumetric DBS device. The volumetric device gave accurate and precise quantifications and the dried capillary blood gave the same results as that of venous blood. image
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinska och farmaceutiska grundvetenskaper -- Farmakologi och toxikologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Basic Medicine -- Pharmacology and Toxicology (hsv//eng)
Nyckelord
- alcohol biomarker
- DBS
- liquid chromatography mass spectrometry
- microsampling
- phosphatidylethanol