Systematisation of spatial uncertainties for comparison between a MR and a CT-based radiotherapy workflow for prostate treatments

Nyholm, Tufve (författare)
Umeå universitet,Radiofysik,Department of Radiation Sciences (Oncology), Umeå University Hospital
Nyberg, Morgan (författare)
Luleå tekniska universitet,Signaler och system
Karlsson, Magnus G (författare)
Umeå universitet,Radiofysik,Department of Radiation Physics, Umeå University Hospital
visa fler...
Karlsson, Mikael (författare)
Umeå universitet,Radiofysik,Radiation physics section, Department of radiation sciences, Umeå University
visa färre...
 (creator_code:org_t)
2009
2009
Engelska.
Ingår i: Radiation Oncology. - 1748-717X. ; 4:54
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: In the present work we compared the spatial uncertainties associated with a MR-based workflow for external radiotherapy of prostate cancer to a standard CT-based workflow. The MR-based workflow relies on target definition and patient positioning based on MR imaging. A solution for patient transport between the MR scanner and the treatment units has been developed. For the CT-based workflow, the target is defined on a MR series but then transferred to a CT study through image registration before treatment planning, and a patient positioning using portal imaging and fiducial markers. Methods: An "open bore" 1.5T MRI scanner, Siemens Espree, has been installed in the radiotherapy department in near proximity to a treatment unit to enable patient transport between the two installations, and hence use the MRI for patient positioning. The spatial uncertainty caused by the transport was added to the uncertainty originating from the target definition process, estimated through a review of the scientific literature. The uncertainty in the CT-based workflow was estimated through a literature review.Results: The systematic uncertainties, affecting all treatment fractions, are reduced from 3-4 mm (ISd) with a CT based workflow to 2-3 mm with a MR based workflow. The main contributing factor to this improvement is the exclusion of registration between MR and CT in the planning phase of the treatment.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Annan medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Other Medical Engineering (hsv//eng)

Nyckelord

Medical Engineering for Healthcare
Medicinsk teknik för hälsovård

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy