Advanced Pressure Compensation in High Accuracy NDIR Sensors for Environmental Studies

Gaynullin, Bakhram, 1967- (författare)
Mittuniversitetet,Institutionen för ingenjörsvetenskap, matematik och ämnesdidaktik (2023-),Senseair AB, Delsbo, Sweden
Hummelgård, Christine (författare)
Senseair AB, Delsbo, Sweden
Mattsson, Claes, 1978- (författare)
Mittuniversitetet,Institutionen för ingenjörsvetenskap, matematik och ämnesdidaktik (2023-)
visa fler...
Thungström, Göran, 1960- (författare)
Mittuniversitetet,Institutionen för ingenjörsvetenskap, matematik och ämnesdidaktik (2023-)
Rödjegård, Henrik (författare)
Senseair AB, Delsbo, Sweden
visa färre...
 (creator_code:org_t)
2023-03-06
2023
Engelska.
Ingår i: Sensors. - : MDPI AG. - 1424-8220. ; 23:5
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Measurements of atmospheric gas concentrations using of NDIR gas sensors requires compensation of ambient pressure variations to achieve reliable result. The extensively used general correction method is based on collecting data for varying pressures for a single reference concentration. This one-dimensional compensation approach is valid for measurements carried out in gas concentrations close to reference concentration but will introduce significant errors for concentrations further away from the calibration point. For applications, requiring high accuracy, collecting, and storing calibration data at several reference concentrations can reduce the error. However, this method will cause higher demands on memory capacity and computational power, which is problematic for cost sensitive applications. We present here an advanced, but practical, algorithm for compensation of environmental pressure variations for relatively low-cost/high resolution NDIR systems. The algorithm consists of a two-dimensional compensation procedure, which widens the valid pressure and concentrations range but with a minimal need to store calibration data, compared to the general one-dimensional compensation method based on a single reference concentration. The implementation of the presented two-dimensional algorithm was verified at two independent concentrations. The results show a reduction in the compensation error from 5.1% and 7.3%, for the one-dimensional method, to −0.02% and 0.83% for the two-dimensional algorithm. In addition, the presented two-dimensional algorithm only requires calibration in four reference gases and the storing of four sets of polynomial coefficients used for calculations. 

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Elektroteknik och elektronik -- Annan elektroteknik och elektronik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Electrical Engineering, Electronic Engineering, Information Engineering -- Other Electrical Engineering, Electronic Engineering, Information Engineering (hsv//eng)

Nyckelord

carbon dioxide
compensation
NDIR
pressure
sensor

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

  • Sensors (Sök värdpublikationen i LIBRIS)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy