Effects of lung tissue characterization in radiotherapy of breast cancer under deep inspiration breath hold when using Monte Carlo dosimetry

  • Artikel/kapitelEngelska2021

Förlag, utgivningsår, omfång ...

  • Associazione Italiana di Fisica Medica,2021
  • printrdacarrier

Nummerbeteckningar

  • LIBRIS-ID:oai:DiVA.org:ri-56926
  • https://urn.kb.se/resolve?urn=urn:nbn:se:ri:diva-56926URI
  • https://doi.org/10.1016/j.ejmp.2021.09.009DOI
  • https://gup.ub.gu.se/publication/310618URI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:ref swepub-contenttype
  • Ämneskategori:art swepub-publicationtype

Anmärkningar

  • Funding details: Västra Götalandsregionen; Funding details: Strålsäkerhetsmyndigheten, SSM; Funding text 1: Financial support of King Gustav the V Jubilee Clinic Foundation, The Swedish Radiation Safety Authority and The Healthcare Committee, Region Västra Götaland are greatly acknowledged.
  • Purpose: To investigate the sensitivity of Monte Carlo (MC) calculated lung dose distributions to lung tissue characterization in external beam radiotherapy of breast cancer under Deep Inspiration Breath Hold (DIBH). Methods: EGSnrc based MC software was employed. Mean lung densities for one hundred patients were analysed. CT number frequency and clinical dose distributions were calculated for 15 patients with mean lung density below 0.14 g/cm3. Lung volume with a pre-defined CT numbers was also considered. Lung tissue was characterized by applying different CT calibrations in the low-density region and air-lung tissue thresholds. Dose impact was estimated by Dose Volume Histogram (DVH) parameters. Results: Mean lung densities below 0.14 g/cm3 were found in 10% of the patients. CT numbers below −960 HU dominated the CT frequency distributions with a high rate of CT numbers at −990 HU. Mass density conversion approach influenced the DVH shape. V4Gy and V8Gy varied by 7% and 5% for the selected patients and by 9% and 3.5% for the pre-defined lung volume. V16Gy and V20Gy, were within 2.5%. Regions above 20 Gy were affected. Variations in air- lung tissue differentiation resulted in DVH parameters within 1%. Threshold at −990 HU was confirmed by the CT number frequency distributions. Conclusions: Lung dose distributions were more sensitive to variations in the CT calibration curve below lung (inhale) density than to air-lung tissue differentiation. Low dose regions were mostly affected. The dosimetry effects were found to be potentially important to 10% of the patients treated under DIBH.

Ämnesord och genrebeteckningar

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Pettersson, NiclasSahlgrenska University Hospital, Sweden (författare)
  • Lindberg, AndreasSahlgrenska University Hospital, Sweden (författare)
  • Swanpalmer, John,1958Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences(Swepub:gu)xswajo (författare)
  • Chakarova, RoumianaGothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för medicinsk strålningsvetenskap,Institute of Clinical Sciences, Department of Medical Radiation Sciences(Swepub:gu)xchakr (författare)
  • RISEUniversity of Gothenburg, Sweden (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Physica medica (Testo stampato): Associazione Italiana di Fisica Medica90, s. 83-901120-17971724-191X

Internetlänk

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy