  • Jakobsen, Stig S. (author)

Failure of total hip implants : metals and metal release in 52 cases

  • Article/chapterEnglish2014

Publisher, publication year, extent ...

  • 2014-07-18
  • Wiley,2014
  • printrdacarrier

Numbers

  • LIBRIS-ID:oai:DiVA.org:su-110742
  • https://urn.kb.se/resolve?urn=urn:nbn:se:su:diva-110742URI
  • https://doi.org/10.1111/cod.12275DOI
  • http://kipublications.ki.se/Default.aspx?queryparsed=id:130083009URI

Supplementary language notes

  • Language:English
  • Summary in:English

Part of subdatabase

Classification

  • Subject category:ref swepub-contenttype
  • Subject category:art swepub-publicationtype

Notes

  • AuthorCount:10;
  • BackgroundThe pathogenesis of total joint replacement failure is multifactorial. One hypothesis suggests that corrosion and wear of alloys result in metal ion release, which may then cause sensitization and even implant failure, owing to the acquired immune reactivity. ObjectivesTo assess cobalt, nickel and chromium(VI) release from, and the metal composition of, failed metal-on-ethylene total hip replacements. Materials/methodsImplant components from 52 revision cases were evaluated with spot tests for free nickel, cobalt, and chromium (VI) ions. Implant composition was determined with X-ray fluorescence spectroscopy, and information on the reason for revision and complications in relation to surgery was collected from the medical charts when possible (72%). For 10 implants, corrosion was further characterized with scanning electron microscopy. ResultsWe detected cobalt release from three of 38 removed femoral heads and from one of 24 femoral stems. Nickel release was detected from one of 24 femoral stems. No chromium(VI) release was detected. ConclusionsWe found that cobalt and nickel were released from some failed total hip arthroplasties, and corrosion was frequently observed. Metal ions and particles corroded from metal-on-polyethylene may play a role in the complex aetiopathology of implant failure.

Subject headings and genre

Added entries (persons, corporate bodies, meetings, titles ...)

  • Liden, CarolaKarolinska Institutet (author)
  • Soballe, Kjeld (author)
  • Johansen, Jeanne D. (author)
  • Menne, Torkil (author)
  • Lundgren, LennartStockholms universitet,Institutionen för tillämpad miljövetenskap (ITM),Karolinska Institutet, Sweden(Swepub:su)lennl (author)
  • Bregnbak, David (author)
  • Moller, Per (author)
  • Jellesen, Morten S. (author)
  • Thyssen, Jacob P. (author)
  • Karolinska InstitutetInstitutionen för tillämpad miljövetenskap (ITM) (creator_code:org_t)

Related titles

  • In:Contact Dermatitis: Wiley71:6, s. 319-3250105-18731600-0536

Internet link

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view