EANM guidelines for PET-CT and PET-MR routine quality control

Koole, Michel (författare)
Nuclear Medicine and Molecular Imaging, Department of Imaging and Pathology, KU Leuven, Belgium
Amstrong, Ian (författare)
Nuclear Medicine, Manchester University NHS Foundation Trust, Manchester, United Kingdom
Krizsan, Aron K. (författare)
ScanoMed Nuclear Medicine Centers, Debrecen, Hungary
visa fler...
Strömvall, Anne (författare)
Umeå universitet,Radiofysik
Visvikis, Dimitris (författare)
LaTIM, INSERM, UMR 1101, Univ Brest, Brest, France
Sattler, Bernhard (författare)
Department of Nuclear Medicine, University Medical Centre Leipzig, Leipzig, Germany
Nekolla, Stephan G. (författare)
Nuklearmedizinische Klinik und Poliklinik, Klinikum rechts der Isar der Technischen Universität München, München, Germany
Dickson, John (författare)
Institute of Nuclear Medicine, University College London Hospital, London, United Kingdom
visa färre...
 (creator_code:org_t)
Elsevier, 2023
2023
Engelska.
Ingår i: Zeitschrift für Medizinische Physik. - : Elsevier. - 0939-3889 .- 1876-4436. ; 33:1, s. 103-113
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We present guidelines by the European Association of Nuclear Medicine (EANM) for routine quality control (QC) of PET-CT and PET-MR systems. These guidelines are partially based on the current EANM guidelines for routine quality control of Nuclear Medicine instrumentation but focus more on the inherent multimodal aspect of the current, state-of-the-art PET-CT and PET-MR scanners. We briefly discuss the regulatory context put forward by the International Electrotechnical Commission (IEC) and European Commission (EC) and consider relevant guidelines and recommendations by other societies and professional organizations. As such, a comprehensive overview of recommended quality control procedures is provided to ensure the optimal operational status of a PET system, integrated with either a CT or MR system. In doing so, we also discuss the rationale of the different tests, advice on the frequency of each test and present the relevant MR and CT tests for an integrated system. In addition, we recommend a scheme of preventive actions to avoid QC tests from drifting out of the predefined range of acceptable performance values such that an optimal performance of the PET system is maintained for routine clinical use.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Guidelines
PET-CT
PET-MR
Quality control

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy